Comparison of Pattern VEP Results Acquired Using CRT and TFT Stimulators in the Clinical Practice

Overview
Journal Doc Ophthalmol
Specialty Ophthalmology
Date 2011 Apr 9
PMID 21475982
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

There are several electrophysiological systems available commercially. Usually, control groups are required to compare their results, due to the differences between display types. Our aim was to examine the differences between CRT and LCD/TFT stimulators used in pattern VEP responses performed according to the ISCEV standards. We also aimed to check different contrast values toward thresholds. In order to obtain more precise results, we intended to measure the intensity and temporal response characteristics of the monitors with photometric methods. To record VEP signals, a Roland RetiPort electrophysiological system was used. The pattern VEP tests were carried out according to ISCEV protocols on a CRT and a TFT monitor consecutively. Achromatic checkerboard pattern was used at three different contrast levels (maximal, 75, 25%) using 1° and 15' check sizes. Both CRT and TFT displays were luminance and contrast matched, according to the gamma functions based on measurements at several DAC values. Monitor-specific luminance parameters were measured by means of spectroradiometric instruments. Temporal differences between the displays' electronic and radiometric signals were measured with a device specifically built for the purpose. We tested six healthy control subjects with visual acuity of at least 20/20. The tests were performed on each subject three times on different days. We found significant temporal differences between the CRT and the LCD monitors at all contrast levels and spatial frequencies. In average, the latency times were 9.0 ms (±3.3 ms) longer with the TFT stimulator. This value is in accordance with the average of the measured TFT input-output temporal difference values (10.1 ± 2.2 ms). According to our findings, measuring the temporal parameters of the TFT monitor with an adequately calibrated measurement setup and correcting the VEP data with the resulting values, the VEP signals obtained with different display types can be transformed to be comparable.

Citing Articles

True pattern-reversal LED stimulator and its comparison to LCD and CRT displays: visual evoked potential study.

Voda P, Kremlacek J, Kordek D, Chutna M, Bezrouk A Sci Rep. 2024; 14(1):4244.

PMID: 38378903 PMC: 10879535. DOI: 10.1038/s41598-024-54776-5.


Assessment of digital light processing (DLP) projector stimulators for visual electrophysiology.

Marmoy O, Thompson D Doc Ophthalmol. 2023; 146(2):151-163.

PMID: 36617339 PMC: 10082110. DOI: 10.1007/s10633-022-09917-4.


Comparison of CRT and LCD monitors for objective estimation of visual acuity using the sweep VEP.

Strasser T, Leinberger D, Hillerkuss D, Zrenner E, Zobor D Doc Ophthalmol. 2022; 145(2):133-145.

PMID: 35788850 PMC: 9470625. DOI: 10.1007/s10633-022-09883-x.


Effect of biological factors on latency of pattern-reversal visual evoked potentials associated with cathode ray tubes and liquid crystal display monitors in normal young subjects.

Ura M, Matsuo M, Yamazaki H, Morita H Doc Ophthalmol. 2021; 143(2):185-192.

PMID: 33847872 DOI: 10.1007/s10633-021-09833-z.


Pattern visual evoked potentials elicited by organic electroluminescence screen.

Matsumoto C, Shinoda K, Matsumoto H, Funada H, Sasaki K, Minoda H Biomed Res Int. 2014; 2014:606951.

PMID: 25197652 PMC: 4147363. DOI: 10.1155/2014/606951.


References
1.
Husain A, Hayes S, Young M, Shah D . Visual evoked potentials with CRT and LCD monitors: when newer is not better. Neurology. 2009; 72(2):162-4. DOI: 10.1212/01.wnl.0000339041.29147.5f. View

2.
Karanjia R, Brunet D, Ten Hove M . Optimization of visual evoked potential (VEP) recording systems. Can J Neurol Sci. 2009; 36(1):89-92. DOI: 10.1017/s0317167100006375. View

3.
Odom J, Bach M, Brigell M, Holder G, McCulloch D, Tormene A . ISCEV standard for clinical visual evoked potentials (2009 update). Doc Ophthalmol. 2009; 120(1):111-9. DOI: 10.1007/s10633-009-9195-4. View

4.
Elze T . Achieving precise display timing in visual neuroscience experiments. J Neurosci Methods. 2010; 191(2):171-9. DOI: 10.1016/j.jneumeth.2010.06.018. View