Quantification of Liver Iron Content with CT-added Value of Dual-energy

Overview
Journal Eur Radiol
Specialty Radiology
Date 2011 Apr 8
PMID 21472472
Citations 28
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To evaluate the value of dual-energy CT (DECT) with use of an iron-specific, three-material decomposition algorithm for the quantification of liver iron content (LIC).

Methods: Thirty-one phantoms containing liver tissue, fat and iron were scanned with dual-source CT using single-energy at 120 kV (SECT) and DECT at 80 kV and 140 kV. Virtual iron concentration (VIC) images derived from an iron-specific, three-material decomposition algorithm and measurements of fat-free and fat-containing phantoms were compared with the LIC and healthy liver tissue.

Results: In the absence of fat significant linear correlations were found between LIC and HU from SECT and VIC (r = 0.984-0.997, p < 0.001) with a detection limit of 145.4 μmol/g LIC for SECT, whereas VIC accurately quantified the lowest LIC of 20 μmol/g dry liver. In the presence of fat, no significant correlation was observed between LIC and SECT, whereas significant correlations were found for VIC. Compared with fat-free phantoms, significant underestimation of LIC was seen for SECT with increasing amounts of fat (all, p ≤ 0.01). On the other hand, similar HU were seen for VIC of fat-containing compared with fat-free phantoms (p > 0.632).

Conclusions: Virtual iron concentration images generated from DECT provide added value for the quantification of LIC by disregarding the confounding effect of the natural variation of healthy liver attenuation and of co-existing liver fat.

Citing Articles

Computed Tomography and Magnetic Resonance Imaging in Liver Iron Overload: From Precise Quantification to Prognosis Assessment.

Zhou X, Jia X, Chen Y, Song B Biomedicines. 2024; 12(11).

PMID: 39595022 PMC: 11592092. DOI: 10.3390/biomedicines12112456.


Hepatic fat quantification in dual-layer computed tomography using a three-material decomposition algorithm.

Demondion E, Ernst O, Louvet A, Robert B, Kafri G, Langzam E Eur Radiol. 2023; 34(6):3708-3718.

PMID: 37955671 DOI: 10.1007/s00330-023-10382-z.


Quantification of bone quality and distribution of the proximal humerus with dual-energy computed tomography.

Yu Q, Yang J, Zhou C, Xu Z, Liu C, Luo Q Quant Imaging Med Surg. 2023; 13(9):5676-5687.

PMID: 37711831 PMC: 10498250. DOI: 10.21037/qims-22-927.


Non-Invasive Imaging Methods to Evaluate Non-Alcoholic Fatty Liver Disease with Fat Quantification: A Review.

Jang W, Song J Diagnostics (Basel). 2023; 13(11).

PMID: 37296703 PMC: 10252963. DOI: 10.3390/diagnostics13111852.


Comparison between Dual-Energy CT and Quantitative Susceptibility Mapping in Assessing Brain Iron Deposition in Parkinson Disease.

Chen Y, Ge M, Kang J, Ding Y, Chen Y, Jia Z AJNR Am J Neuroradiol. 2023; 44(4):410-416.

PMID: 36958800 PMC: 10084894. DOI: 10.3174/ajnr.A7822.


References
1.
Bonkovsky H, Lambrecht R, Shan Y . Iron as a co-morbid factor in nonhemochromatotic liver disease. Alcohol. 2003; 30(2):137-44. DOI: 10.1016/s0741-8329(03)00127-7. View

2.
GOLDBERG H, Cann C, Moss A, OHTO M, Brito A, Federle M . Noninvasive quantitation of liver iron in dogs with hemochromatosis using dual-energy CT scanning. Invest Radiol. 1982; 17(4):375-80. DOI: 10.1097/00004424-198207000-00013. View

3.
Gandon Y, Guyader D, Heautot J, Reda M, Yaouanq J, Buhe T . Hemochromatosis: diagnosis and quantification of liver iron with gradient-echo MR imaging. Radiology. 1994; 193(2):533-8. DOI: 10.1148/radiology.193.2.7972774. View

4.
Boll D, Merkle E . Diffuse liver disease: strategies for hepatic CT and MR imaging. Radiographics. 2009; 29(6):1591-614. DOI: 10.1148/rg.296095513. View

5.
Karlo C, Lauber A, Gotti R, Baumuller S, Stolzmann P, Scheffel H . Dual-energy CT with tin filter technology for the discrimination of renal lesion proxies containing blood, protein, and contrast-agent. An experimental phantom study. Eur Radiol. 2010; 21(2):385-92. DOI: 10.1007/s00330-010-1926-0. View