» Articles » PMID: 21442578

The C-Port XV® Vascular Anastomosis System: Results from an Animal Trial

Overview
Date 2011 Mar 29
PMID 21442578
Authors
Affiliations
Soon will be listed here.
Abstract

Background: In this study, facilitated anastomosis using an anastomotic device was compared to conventional hand-sewn (HS) vascular anastomosis in an animal model.

Methods: A pig carotid bypass model was employed. C-Port xV® (xV) and HS anastomoses were compared by evaluating intraoperative performance, midterm graft patency, and histology.

Results: All animals survived; none developed early/late neurological deficits. Mean graft blood flow was comparable between groups (HS group: 161 ± 61 ml/min; xV group: 143 ± 44 ml/min). All anastomoses were patent at necropsy (at 111 ± 6 postoperative days). Histologically, no significant inflammation was found around the fasteners or in the vessel wall. Neointimal overgrowth on the lumen surface appeared organized and covered with endothelium. There was no adherence of fibrin, platelets, or inflammatory cells to the surface. The neointimal tissue appeared normal without any inflammation, hemorrhage, calcification, or necrosis.