Biomechanical Analysis of Occipitocervical Stability Afforded by Three Fixation Techniques

Overview
Journal Spine J
Specialty Orthopedics
Date 2011 Mar 8
PMID 21377608
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Background Context: Occipital condyle screws appear to be a novel technique that demands biomechanical consideration. It has the potential to achieve fixation anterior to the axis of rotation while offering a point of fixation in line with the C1/C2 screws.

Purpose: To compare the segmental stability and range of motion (ROM) of standard occipitocervical (OC) screw/rod and plate constructs versus a new technique that incorporates occipital condyle fixation.

Study Design: Human cadaveric biomechanical analysis.

Methods: After intact analysis, 10 fresh-frozen human cadaveric OC spine specimens were instrumented bilaterally with C1 lateral mass screws and C2 pedicle screws. Additional occipital instrumentation was tested in random order under the following conditions: standard occipitocervical plate/rod system (Vertex Max; Medtronic, Inc., Minneapolis, MN, USA); occipital condyle screws alone; and occipital condyle screws with the addition of an eyelet screw placed into the occiput bilaterally. After nondestructive ROM testing, specimens were evaluated under computed tomography (CT) and underwent destructive forward flexion failure comparing Group 1 to Group 3.

Results: There was no significant difference in OC (Occiput-C1) axial rotation and flexion/extension ROM between the standard occipitocervical plate/rod system (Group 1) and the occipital condyle screws with one eyelet screw bilaterally (Group 3). Furthermore, the occipital condyle screws alone (Group 2) did allow significantly more flexion/extension compared with Group 1. Interestingly, the two groups with occipital condyle screws (Groups 2 and 3) had significantly less lateral bending compared with Group 1. During CT analysis, the mean occipital condyle width was 10.8 mm (range, 9.1-12.7 mm), and the mean condylar length was 24.3 mm (range, 20.2-28.5). On destructive testing, there was no significant difference in forward flexion failure between Groups 1 and 3.

Conclusions: With instrumentation across the mobile OC junction, our results indicate that similar stability can be achieved with occipital condyle screws/eyelet screws compared with the standard occipitocervical plate/rod system.

Citing Articles

Optimal trajectory for the anterior occipital condyle screw.

Lou Z, Zhu J, Jiang K, Xia S, Chen S, Zhu J J Int Med Res. 2025; 53(3):3000605251325677.

PMID: 40073407 PMC: 11905021. DOI: 10.1177/03000605251325677.


Headache disorders in patients with Ehlers-Danlos syndromes and hypermobility spectrum disorders.

Mehta D, Simmonds L, Hakim A, Matharu M Front Neurol. 2024; 15:1460352.

PMID: 39582682 PMC: 11581963. DOI: 10.3389/fneur.2024.1460352.


The Feasibility of Condylar Screws for Occipitocervical Fusion in Arabs: Computed Tomography-Based Morphometric Study.

Marwan Y, Jarragh A, Algarni N, Sheikh M, Alsaeed O Global Spine J. 2024; :21925682241268093.

PMID: 39095957 PMC: 11571779. DOI: 10.1177/21925682241268093.


A Novel Technique for Basilar Invagination Treatment in a Patient with Klippel-Feil Syndrome: A Clinical Example and Brief Literature Review.

Tanaka M, Al Askar A, Kumawat C, Arataki S, Komatsubara T, Taoka T Medicina (Kaunas). 2024; 60(4).

PMID: 38674263 PMC: 11051808. DOI: 10.3390/medicina60040616.


The modified occipital condyle screw: A quantitative anatomic study investigating the feasibility of a novel instrumented fixation technique for craniocervical fusion.

Dietz N, Gruter B, Haegler J, Cho S, Tubbs R, Spiessberger A J Craniovertebr Junction Spine. 2023; 14(2):175-180.

PMID: 37448506 PMC: 10336899. DOI: 10.4103/jcvjs.jcvjs_27_23.