An Algorithm for Efficient Metal Artifact Reductions in Permanent Seed

Overview
Journal Med Phys
Specialty Biophysics
Date 2011 Mar 3
PMID 21361174
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: In permanent seed implants, 60 to more than 100 small metal capsules are inserted in the prostate, creating artifacts in x-ray computed tomography (CT) imaging. The goal of this work is to develop an automatic method for metal artifact reduction (MAR) from small objects such as brachytherapy seeds for clinical applications.

Methods: The approach for MAR is based on the interpolation of missing projections by directly using raw helical CT data (sinogram). First, an initial image is reconstructed from the raw CT data. Then, the metal objects segmented from the reconstructed image are reprojected back into the sinogram space to produce a metal-only sinogram. The Steger method is used to determine precisely the position and edges of the seed traces in the raw CT data. By combining the use of Steger detection and reprojections, the missing projections are detected and replaced by interpolation of non-missing neighboring projections.

Results: In both phantom experiments and patient studies, the missing projections have been detected successfully and the artifacts caused by metallic objects have been substantially reduced. The performance of the algorithm has been quantified by comparing the uniformity between the uncorrected and the corrected phantom images. The results of the artifact reduction algorithm are indistinguishable from the true background value.

Conclusions: An efficient algorithm for MAR in seed brachytherapy was developed. The test results obtained using raw helical CT data for both phantom and clinical cases have demonstrated that the proposed MAR method is capable of accurately detecting and correcting artifacts caused by a large number of very small metal objects (seeds) in sinogram space. This should enable a more accurate use of advanced brachytherapy dose calculations, such as Monte Carlo simulations.

Citing Articles

Automatic recognition and analysis of metal streak artifacts in head and neck computed tomography for radiomics modeling.

Wei L, Rosen B, Vallieres M, Chotchutipan T, Mierzwa M, Eisbruch A Phys Imaging Radiat Oncol. 2021; 10:49-54.

PMID: 33458268 PMC: 7807651. DOI: 10.1016/j.phro.2019.05.001.


Image-based Metal Artifact Reduction in X-ray Computed Tomography utilizing Local Anatomical Similarity.

Dong X, Yang X, Rosenfield J, Elder E, Dhabaan A Proc SPIE Int Soc Opt Eng. 2019; 10132.

PMID: 31456599 PMC: 6711614. DOI: 10.1117/12.2255083.


The effects of the orthopedic metal artifact reduction (O-MAR) algorithm on contouring and dosimetry of head and neck radiotherapy patients.

Sillanpaa J, Lovelock M, Mueller B Med Dosim. 2019; 45(1):92-96.

PMID: 31375297 PMC: 6989373. DOI: 10.1016/j.meddos.2019.07.003.


A metal artifact reduction method for a dental CT based on adaptive local thresholding and prior image generation.

Hegazy M, Cho M, Lee S Biomed Eng Online. 2016; 15(1):119.

PMID: 27814775 PMC: 5097357. DOI: 10.1186/s12938-016-0240-8.


Spectral CT with monochromatic imaging and metal artifacts reduction software for artifacts reduction of ¹²⁵I radioactive seeds in liver brachytherapy.

Yang Q, Peng S, Wu J, Ban X, He M, Xie C Jpn J Radiol. 2015; 33(11):694-705.

PMID: 26456321 DOI: 10.1007/s11604-015-0482-6.