» Articles » PMID: 21241912

Automated Keratometry in Routine Cataract Surgery: Comparison of Scheimpflug and Conventional Values

Overview
Specialty Ophthalmology
Date 2011 Jan 19
PMID 21241912
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare the mean keratometry (K) readings obtained with a conventional automated keratometer (IOLMaster) and a Scheimpflug keratometer (Pentacam) in eyes having preoperative assessment for routine cataract surgery.

Setting: Epsom and St. Helier University Hospitals, London, United Kingdom.

Design: Evaluation of diagnostic technology.

Methods: Mean K values were obtained with the conventional and Scheimpflug keratometers. The following Scheimpflug readings were evaluated: anterior K, true net power, and Holladay equivalent K measured at 1.0 to 7.0 mm corneal diameters. Mean readings for each type of keratometry were compared. Bland-Altman plots were used to determine the 95% limits of agreement (LoA) for the conventional and Scheimpflug keratometers.

Results: The mean conventional K was statistically significantly greater than the mean Scheimpflug K for true net power and equivalent K at 1.0 mm, 2.0 mm, 3.0 mm, and 4.0 mm corneal diameters. The mean conventional K was significantly less than the equivalent K at 5.0 mm, 6.0 mm, and 7.0 mm. The smallest mean difference was for equivalent K at 4.5 mm (0.02 diopters [D]). The smallest 95% LoA were -0.68 to 1.16 D (equivalent K at 5.0 mm) and -0.91 to 0.95 D (equivalent K at 4.5 mm).

Conclusions: Overall, the equivalent K at 4.5 mm had the closest match with the conventional K values. The degree of interdevice variability with the conventional keratometer was lowest for the equivalent K at 4.5 mm and 5.0 mm, although this variability would be sufficient to influence intraocular lens power selection.

Citing Articles

Comparison of total corneal power measurements obtained with different devices after myopic keratorefractive surgery.

Wang Z, Song Y, Yang W, Liu Q, Li Y, Cui R Int J Ophthalmol. 2023; 16(7):1117-1122.

PMID: 37465503 PMC: 10333248. DOI: 10.18240/ijo.2023.07.17.


Comparing Standard Keratometry and Total Keratometry Before and After Myopic Corneal Refractive Surgery With a Swept-Source OCT Biometer.

Wang Z, Song Y, Yang W, Li D, Chen W, Zhao Q Front Med (Lausanne). 2022; 9:928027.

PMID: 35903314 PMC: 9318577. DOI: 10.3389/fmed.2022.928027.


Comparison of Total Corneal Astigmatism between IOLMaster and Pentacam.

Li X, Cao X, Bao Y Biomed Res Int. 2022; 2022:9236006.

PMID: 35845936 PMC: 9286877. DOI: 10.1155/2022/9236006.


Comparison of IOL-Master 700 and IOL-Master 500 biometers in ocular biological parameters of adolescents.

Shi Q, Wang G, Cheng Y, Pei C Int J Ophthalmol. 2021; 14(7):1013-1017.

PMID: 34282385 PMC: 8243180. DOI: 10.18240/ijo.2021.07.08.


Evaluation and comparison of a novel Scheimpflug-based optical biometer with standard partial coherence interferometry for biometry and intraocular lens power calculation.

Wang Z, Yang W, Li D, Chen W, Zhao Q, Li Y Exp Ther Med. 2021; 21(4):326.

PMID: 33732299 PMC: 7903467. DOI: 10.3892/etm.2021.9757.