Multicenter Crossover Study of Automated Control of Inspired Oxygen in Ventilated Preterm Infants

Overview
Journal Pediatrics
Specialty Pediatrics
Date 2010 Dec 29
PMID 21187305
Citations 39
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To determine the efficacy and safety of automated adjustment of the fraction of inspired oxygen (Fio(2)) adjustment in maintaining arterial oxygen saturation (Spo(2)) within an intended range for mechanically ventilated preterm infants with frequent episodes of decreased Spo(2).

Methods: Thirty-two infants (gestational age [median and interquartile range]: 25 weeks [24-27 weeks]; age: 27 days [17-36 days]) were studied during 2 consecutive 24-hour periods, one with Fio(2) adjusted by clinical staff members (manual) and the other by an automated system (automated), in random sequence.

Results: Time with Spo(2) within the intended range (87%-93%) increased significantly during the automated period, compared with the manual period (40% ± 14% vs 32% ± 13% [mean ± SD]). Times with Spo(2) of >93% or >98% were significantly reduced during the automated period (21% ± 20% vs 37% ± 12% and 0.7% vs 5.6% [interquartile ranges: 0.1%-7.2% and 2.7%-11.2%], respectively). Time with Spo(2) of <87% increased significantly during the automated period (32% ± 12% vs 23% ± 9%), with more-frequent episodes with Spo(2) between 80% and 86%, whereas times with Spo(2) of <80% or <75% did not differ between periods. Hourly median Fio(2) values throughout the automated period were lower and there were substantially fewer manual Fio(2) changes (10 ± 9 vs 112 ± 59 changes per 24 hours; P < .001), compared with the manual period.

Conclusions: In infants with fluctuations in Spo(2), automated Fio(2) adjustment improved maintenance of the intended Spo(2) range led to reduced time with high Spo(2) and more-frequent episodes with Spo(2) between 80% and 86%.

Citing Articles

Statistical analysis plan for the FiO-C trial: effects of closed-loop automatic control of the inspiratory fraction of oxygen (FiO-C) on outcomes of extremely preterm infants-a randomized-controlled parallel group multicentre trial for safety and....

Konig J, Stauch A, Engel C, Urschitz M, Franz A Trials. 2024; 25(1):756.

PMID: 39533330 PMC: 11559155. DOI: 10.1186/s13063-024-08615-7.


Oxygen in the neonatal ICU: a complicated history and where are we now?.

Mamidi R, McEvoy C Front Pediatr. 2024; 12:1371710.

PMID: 38751747 PMC: 11094359. DOI: 10.3389/fped.2024.1371710.


Effects of closed loop ventilation on ventilator settings, patient outcomes and ICU staff workloads - a systematic review.

Goossen R, Schultz M, Tschernko E, Chew M, Robba C, Paulus F Eur J Anaesthesiol. 2024; 41(6):438-446.

PMID: 38385449 PMC: 11064903. DOI: 10.1097/EJA.0000000000001972.


Extended CPAP or low-flow nasal cannula for intermittent hypoxaemia in preterm infants: a 24-hour randomised clinical trial.

Yazdi S, Carlo W, Nakhmani A, Boateng E, Aban I, Ambalavanan N Arch Dis Child Fetal Neonatal Ed. 2024; 109(5):557-561.

PMID: 38365446 PMC: 11327380. DOI: 10.1136/archdischild-2023-326605.


Closed-loop anesthesia: foundations and applications in contemporary perioperative medicine.

Coeckelenbergh S, Boelefahr S, Alexander B, Perrin L, Rinehart J, Joosten A J Clin Monit Comput. 2024; 38(2):487-504.

PMID: 38184504 DOI: 10.1007/s10877-023-01111-4.