» Articles » PMID: 20924265

A New Dual Port Cutter System for Vitrectomy Surgery

Overview
Journal Retina
Date 2010 Oct 7
PMID 20924265
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate and compare the performance of several designs of 20-gauge dual port (DP) vitreous cutter tips with a standard 20-gauge single port tip.

Methods: Custom 20-gauge pneumatic vitreous DP cutter tips with different sizes and port positions were evaluated through porcine vitreous and water flow rates. Five designed and fabricated DP cutter tips were compared with a normal single port control tip and evaluated by the measurement of water and porcine vitreous flow rates, and surgical examination in enucleated porcine eyes.

Results: Some DP tips approached a maximum vitreous and water flow rates, removing water and vitreous faster than the normal control tip. With reference to surgical evaluation, some DP tips performed better than the single port tip for bulk vitrectomy, but none shaved the vitreous base more effectively.

Conclusion: The DP cutter system has the potential to increase the flow rates depending on the size and position of the extra port. In the future, the DP cutter may allow the surgeon to perform bulk vitrectomy more efficiently.

Citing Articles

Adverse Events Associated With 3 Vitrectomy Platforms Reported to the US FDA MAUDE Database.

Hwang H, Botsford B, Lenis T, Sadiq M, Mahrous A, Orlin A J Vitreoretin Dis. 2024; 8(5):546-553.

PMID: 39318983 PMC: 11418686. DOI: 10.1177/24741264241264356.


Performance, safety and efficiency comparison between 10,000 and 5000 cuts per minute vitrectomy using a 25G cutter: a prospective randomized controlled study.

Fung N, Mak A, Brelen M, Tsang C, Mohamed S, Lam W Int J Retina Vitreous. 2023; 9(1):15.

PMID: 36945036 PMC: 10032027. DOI: 10.1186/s40942-023-00452-1.


In Vitro Comparison of Various High-Speed Vitrectomy Machines Using Dual Blade Cutters.

Tayyab H, Jamil S, Hashmi S Cureus. 2021; 13(5):e15021.

PMID: 34136317 PMC: 8197787. DOI: 10.7759/cureus.15021.


Comparison of modern high-speed vitrectomy systems and the advantages of using dual-bladed probes.

Oravecz R, Uthoff D, Schrage N, Dutescu R Int J Retina Vitreous. 2021; 7(1):8.

PMID: 33468260 PMC: 7814550. DOI: 10.1186/s40942-020-00277-2.


Performance analysis of a new hypersonic vitrector system.

Stanga P, Pastor-Idoate S, Zambrano I, Carlin P, McLeod D PLoS One. 2017; 12(6):e0178462.

PMID: 28586375 PMC: 5460816. DOI: 10.1371/journal.pone.0178462.