» Articles » PMID: 20859294

Susceptibility of Tmax to Tracer Delay on Perfusion Analysis: Quantitative Evaluation of Various Deconvolution Algorithms Using Digital Phantoms

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

The time-to-maximum of the tissue residue function (T(max)) perfusion index has proven very predictive of infarct growth in large clinical trials, yet its dependency on simple tracer delays remains unknown. Here, we determine the dependency of computed tomography (CT) perfusion (CTP) T(max) estimates on tracer delay using a range of deconvolution techniques and digital phantoms. Digital phantom data sets simulating the tracer delay were created from CTP data of six healthy individuals, in which time frames of the left cerebral hemisphere were shifted forward and backward by up to ±5 seconds. These phantoms were postprocessed with three common singular value decomposition (SVD) deconvolution algorithms-standard SVD (sSVD), block-circulant SVD (bSVD), and delay-corrected SVD (dSVD)-with an arterial input function (AIF) obtained from the right middle cerebral artery (MCA). The T(max) values of the left hemisphere were compared among different tracer delays and algorithms by a region of interest-based analysis. The T(max) values by sSVD were positively correlated with 'positive shifts' but unchanged with 'negative shifts,' those by bSVD had an excellent positive linear correlation with both positive and negative shifts, and those by dSVD were relatively constant, although slightly increased with the positive shifts. The T(max) is a parameter highly dependent on tracer delays and deconvolution algorithm.

Citing Articles

Carotid Dolichoarteriopathy (Elongation) of the Carotid Arteries in Patients with Ischemic Stroke Anamnesis.

Golovin D, Rostovtseva T, Kudryavtsev Y, Berdalin A, Lelyuk S, Lelyuk V Biomedicines. 2023; 11(10).

PMID: 37893124 PMC: 10604229. DOI: 10.3390/biomedicines11102751.


CT Perfusion as a Predictor of the Final Infarct Volume in Patients with Tandem Occlusion.

Lacidogna G, Pitocchi F, Mascolo A, Marrama F, DAgostino F, Rocco A J Pers Med. 2023; 13(2).

PMID: 36836576 PMC: 9964425. DOI: 10.3390/jpm13020342.


Penumbra Detection With Oxygen Extraction Fraction Using Magnetic Susceptibility in Patients With Acute Ischemic Stroke.

Uchida Y, Kan H, Inoue H, Oomura M, Shibata H, Kano Y Front Neurol. 2022; 13:752450.

PMID: 35222239 PMC: 8873150. DOI: 10.3389/fneur.2022.752450.


Different Grades of Collateral Circulation for Evaluating Cerebral Hemodynamic Status in Carotid Artery Stenosis.

Fan W, Shi W, Rong J, Guo W, Lu S, Tan J J Healthc Eng. 2022; 2022:8484977.

PMID: 35154622 PMC: 8828319. DOI: 10.1155/2022/8484977.


Normative distribution of posterior circulation tissue time-to-maximum: Effects of anatomic variation, tracer kinetics, and implications for patient selection in posterior circulation ischemic stroke.

Goldman-Yassen A, Straka M, Uhouse M, Dehkharghani S J Cereb Blood Flow Metab. 2021; 41(8):1912-1923.

PMID: 33444095 PMC: 8327115. DOI: 10.1177/0271678X20982395.


References
1.
Prosser J, Butcher K, Allport L, Parsons M, MacGregor L, Desmond P . Clinical-diffusion mismatch predicts the putative penumbra with high specificity. Stroke. 2005; 36(8):1700-4. DOI: 10.1161/01.STR.0000173407.40773.17. View

2.
Mlynash M, Eyngorn I, Bammer R, Moseley M, Tong D . Automated method for generating the arterial input function on perfusion-weighted MR imaging: validation in patients with stroke. AJNR Am J Neuroradiol. 2005; 26(6):1479-86. PMC: 8149072. View

3.
Calamante F, Gadian D, Connelly A . Delay and dispersion effects in dynamic susceptibility contrast MRI: simulations using singular value decomposition. Magn Reson Med. 2000; 44(3):466-73. DOI: 10.1002/1522-2594(200009)44:3<466::aid-mrm18>3.0.co;2-m. View

4.
Albers G, Thijs V, Wechsler L, Kemp S, Schlaug G, Skalabrin E . Magnetic resonance imaging profiles predict clinical response to early reperfusion: the diffusion and perfusion imaging evaluation for understanding stroke evolution (DEFUSE) study. Ann Neurol. 2006; 60(5):508-517. DOI: 10.1002/ana.20976. View

5.
Calamante F, Christensen S, Desmond P, Ostergaard L, Davis S, Connelly A . The physiological significance of the time-to-maximum (Tmax) parameter in perfusion MRI. Stroke. 2010; 41(6):1169-74. DOI: 10.1161/STROKEAHA.110.580670. View