Visceral Arterial Compromise During Intra-aortic Balloon Counterpulsation Therapy

Overview
Journal Circulation
Date 2010 Sep 15
PMID 20837932
Citations 23
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Intraaortic balloon pump (IABP) therapy is a widely used method of circulatory support. Based on frequent findings of balloon malposition with visceral arterial compromise on computerized tomographic (CT) imaging studies, we conducted a systematic review of cardiac surgical IABP patients with available CT scans to determine reasons, incidence, and clinical relevance of malposition.

Methods And Results: From January 2007 to March 2009, a total of 621 of 7756 cardiac surgical patients (8.0%) received perioperative IABP support, of whom 63 (10.1%) received a thoracoabdominal CT scan during IABP support. Proximal and distal balloon positions were analyzed. The anatomic distance from the left subclavian artery to celiac trunk and aortic transverse diameter were measured in all patients and compared with implanted balloon dimensions. Mean age was 67.1±11.9 years; 33.3% were female, and height was 169±9 cm. Based on radiography, proximal balloon position was correct in 96.8% but only appropriate in 38.1% based on CT. In 61 of 63 patients, compromise of at least 1 visceral artery was found: celiac trunk, 96.8%; superior mesenteric artery, 87.3%; and renal arteries, 66.7%. Left subclavian artery to celiac trunk distance was 241±23 mm, and balloon length was 248±17 mm and corresponded to an anatomic to balloon length mismatch in 68.2%. Spinal deformations were found in 42.9%. Laparotomy for mesenteric ischemia was required in 23.8%. Hospital mortality rate was 60.3%.

Conclusions: IABP malposition was commonly identified by CT. Reasons included incorrect proximal balloon position as well as an anatomic-to-balloon length mismatch. Thus, shorter than recommended balloon sizes and better positioning strategies had to be considered.

Citing Articles

Mechanical Circulatory Support for Complex High-risk Percutaneous Coronary Intervention.

Truesdell A, Davies R, Eltelbany M, Megaly M, Rosner C, Cilia L US Cardiol. 2024; 17:e03.

PMID: 39493941 PMC: 11526491. DOI: 10.15420/usc.2022.26.


Role of Mechanical Circulatory Support in Complex High-Risk and Indicated Percutaneous Coronary Intervention: Current Indications, Device Options, and Potential Complications.

Di Muro F, Bellino M, Esposito L, Attisano T, Meucci F, Mattesini A J Clin Med. 2024; 13(16).

PMID: 39201073 PMC: 11355104. DOI: 10.3390/jcm13164931.


CT and chest radiography in evaluation of mechanical circulatory support devices for acute heart failure.

Chaturvedi A, Rotman Y, Hoang T, Jew G, Mandalapu A, Narins C Insights Imaging. 2023; 14(1):122.

PMID: 37454301 PMC: 10350447. DOI: 10.1186/s13244-023-01469-8.


Pressure-strain loops unveil haemodynamics behind mechanical circulatory support systems.

Landra F, Mandoli G, Sciaccaluga C, Gallone G, Bruno F, Fusi C ESC Heart Fail. 2023; 10(4):2607-2620.

PMID: 37345220 PMC: 10375099. DOI: 10.1002/ehf2.14339.


Radiology of Intra-Aortic Balloon Pump Catheters.

Rhodes N, Johnson T, Boyum J, Khandelwal A, Howell B, Froemming A Radiol Cardiothorac Imaging. 2022; 4(2):e210120.

PMID: 35506140 PMC: 9059237. DOI: 10.1148/ryct.210120.