Fractographic Analysis of a Dental Zirconia Framework: a Case Study on Design Issues

Overview
Publisher Elsevier
Date 2010 Sep 10
PMID 20826369
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Fractographic analysis of clinically failed dental ceramics can provide insights as to the failure origin and related mechanisms. One anterior 6-unit all-ceramic zirconia fixed partial denture (FPD) (Cercon) has been clinically recovered and examined using qualitative fractography. The purpose was to identify the fracture origin and to state the reasons for failure. The recovered parts of the zirconia FPD were microscopically examined to identify classic fractographic patterns such as arrest lines, hackle, twist hackle and wake hackle. The direction of crack propagation was mapped and interpreted back to the origin of failure at the interface of the occlusal-palatal tip of the core and the veneering ceramic. An inappropriate core drop design favoring localized stress concentration combined with a pore cluster in the veneering ceramic at the core tip interface were the reasons for this premature through-the-core thickness failure.

Citing Articles

Initial damage and failure load of zirconia-ceramic and metal-ceramic posterior cantilever fixed partial dentures.

Bomicke W, Boisseree P, Rammelsberg P, Rues S Clin Oral Investig. 2024; 28(1):94.

PMID: 38221600 PMC: 10788321. DOI: 10.1007/s00784-024-05501-z.


Increasing dental zirconia micro-retentive aspect through ultra-short pulsed laser microstructuring: study on flexural strength and crystal phase characterization.

Garofalo S, Wehner M, Dohrn A, Bilandzic M, Roos C, Wierichs R Clin Oral Investig. 2021; 26(1):939-955.

PMID: 34402980 PMC: 8791917. DOI: 10.1007/s00784-021-04077-2.


Technical and Biological Complications of Screw-Retained (CAD/CAM) Monolithic and Partial Veneer Zirconia for Fixed Dental Prostheses on Posterior Implants Using a Digital Workflow: A 3-Year Cross-Sectional Retrospective Study.

De Angelis P, Gasparini G, Camodeca F, De Angelis S, Liguori M, Rella E Biomed Res Int. 2021; 2021:5581435.

PMID: 34307657 PMC: 8279852. DOI: 10.1155/2021/5581435.


Effects of Finish Line Design and Fatigue Cyclic Loading on Phase Transformation of Zirconia Dental Ceramics: A Qualitative Micro-Raman Spectroscopic Analysis.

Sorrentino R, Navarra C, Di Lenarda R, Breschi L, Zarone F, Cadenaro M Materials (Basel). 2019; 12(6).

PMID: 30875849 PMC: 6470560. DOI: 10.3390/ma12060863.


Fracture resistance of three-unit zirconia fixed partial denture with modified framework.

Partiyan A, Osman E, Rayyan M, Aboushelib M, Ibrahim A, Jimbo R Odontology. 2016; 105(1):62-67.

PMID: 27083582 DOI: 10.1007/s10266-016-0242-9.


References
1.
Lohbauer U, Petschelt A, Greil P . Lifetime prediction of CAD/CAM dental ceramics. J Biomed Mater Res. 2002; 63(6):780-5. DOI: 10.1002/jbm.10468. View

2.
Tinschert J, Schulze K, Natt G, Latzke P, Heussen N, Spiekermann H . Clinical behavior of zirconia-based fixed partial dentures made of DC-Zirkon: 3-year results. Int J Prosthodont. 2008; 21(3):217-22. View

3.
Vult von Steyern P, Carlson P, Nilner K . All-ceramic fixed partial dentures designed according to the DC-Zirkon technique. A 2-year clinical study. J Oral Rehabil. 2005; 32(3):180-7. DOI: 10.1111/j.1365-2842.2004.01437.x. View

4.
Aboushelib M, Feilzer A, Kleverlaan C . Bridging the gap between clinical failure and laboratory fracture strength tests using a fractographic approach. Dent Mater. 2008; 25(3):383-91. DOI: 10.1016/j.dental.2008.09.001. View

5.
Ernst C, Cohnen U, Stender E, Willershausen B . In vitro retentive strength of zirconium oxide ceramic crowns using different luting agents. J Prosthet Dent. 2005; 93(6):551-8. DOI: 10.1016/j.prosdent.2005.04.011. View