Multi-projection Fluorescence Optical Tomography Using a Handheld-probe-based Optical Imager: Phantom Studies
Overview
Affiliations
A handheld-probe-based optical imager has recently been developed toward three-dimensional tomography. In this study, the improvement of target depth recovery was demonstrated using a multi-projection technique on large slab phantoms using 0.45 cc fluorescing target(s) (with 1:0 contrast ratio) of 1.5 to 2.5 cm deep. Tomographic results using single- and multi- (here dual) projection measurements (with and without a priori information of target location) were compared. In all experimental cases, the use of multi-projection measurements along with a priori information recovered target depth and location closer to their true values, demonstrating its applicability for clinical translation.
Erickson S, Martinez S, DeCerce J, Romero A, Caldera L, Godavarty A Phys Med Biol. 2013; 58(5):1563-79.
PMID: 23417060 PMC: 3619223. DOI: 10.1088/0031-9155/58/5/1563.