Patients' Evaluation of Nocturnal Hypoglycaemia with GlucoDay Continuous Glucose Monitoring in Paediatric Patients
Overview
Authors
Affiliations
A study was conducted to evaluate the accuracy of GlucoDay (A. Menarini Diagnostics) during 48 h of continuous glucose monitoring (CGMS) in type 1 diabetic adolescents and use this novel approach to assess otherwise ignored nocturnal hypoglycaemias, in relationship to intermediate-acting insulin administration timing. Twenty type 1 diabetic adolescents with poor metabolic control were selected from our out-patient department. Equal doses of intermediate insulin were administered at 19:00 and at 22:00 of the first and second night of the study, respectively. Correlation coefficient between GlucoDay and standard glucometer was 0.94; 98.3% of data fall in the A + B area of Error Grid Analysis and 1.7% in the D area. The mean error was 13.9% overall and 16.4% with blood glucose values (BGV) <75 mg/dl. The accuracy, ±15 mg/dl, was 82% for BGV <75 mg/dl and 74% for BGV >75 mg/dl. The CGMS discovered nocturnal hypoglycaemia (NH) in 12/18 patients, but no severe hypoglycaemia. During the first night, 8 asymptomatic NH episodes were found with BGV <60 mg/dl and 12 with BGV <80. During the second night, 4 asymptomatic NH episodes with BGV <60 mg/dl and 5 with BGV <80 were found. Furthermore, during the second night, the mean duration of BGV <126 mg/dl was lower than in the first night. GlucoDay is a reliable device for CGMS in paediatric patients and able to determine asymptomatic NH. Bedtime insulin injections provided safer glycaemic profiles and a lower percentage of hypoglycaemic events, representing a safer insulin administration scheme.
Freckmann G, Eichenlaub M, Waldenmaier D, Pleus S, Wehrstedt S, Haug C J Diabetes Sci Technol. 2023; 17(6):1506-1526.
PMID: 37599389 PMC: 10658695. DOI: 10.1177/19322968231190941.
Edge J, Acerini C, Campbell F, Hamilton-Shield J, Moudiotis C, Rahman S Arch Dis Child. 2017; 102(6):543-549.
PMID: 28137708 PMC: 5466923. DOI: 10.1136/archdischild-2016-311530.
Susceptibility of interstitial continuous glucose monitor performance to sleeping position.
Mensh B, Wisniewski N, Neil B, Burnett D J Diabetes Sci Technol. 2013; 7(4):863-70.
PMID: 23911167 PMC: 3879750. DOI: 10.1177/193229681300700408.
Sparacino G, Zanon M, Facchinetti A, Zecchin C, Maran A, Cobelli C Sensors (Basel). 2012; 12(10):13753-80.
PMID: 23202020 PMC: 3545591. DOI: 10.3390/s121013753.