» Articles » PMID: 20624255

Diagnostic Accuracy of a Questionnaire and Simple Home Monitoring Device in Detecting Obstructive Sleep Apnoea in a Chinese Population at High Cardiovascular Risk

Overview
Journal Respirology
Specialty Pulmonary Medicine
Date 2010 Jul 14
PMID 20624255
Citations 36
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Objective: OSA is a common condition associated with cardiovascular (CV) morbidity. It remains underdiagnosed globally in part due to the limited availability and technical requirements of polysomnography (PSG). The aim of this study was to test the accuracy of two simple methods for diagnosing OSA.

Methods: Consecutive subjects identified from a community register with high CV risk were invited to complete the Berlin Sleep Questionnaire and undergo simultaneous, home, overnight PSG and ApneaLink device oximetry and nasal pressure recordings. The relative accuracies of the Berlin Questionnaire, oximetry and nasal pressure results in diagnosing PSG-defined moderate-severe OSA were assessed.

Results: Of 257 eligible high CV risk subjects enrolled, 190 completed sleep studies and 143 subjects' studies were of sufficient quality to include in final analyses. Moderate-severe OSA was confirmed in 43% of subjects. The Berlin Questionnaire had low overall diagnostic accuracy in this population. However, ApneaLink recordings of oximetry and nasal pressure areas had high diagnostic utility with areas under the receiver operating characteristic curves of 0.933 and 0.933, respectively. At optimal diagnostic thresholds, oximetry and nasal pressure measurements had similar sensitivity (84% vs 86%) and specificity (84% vs 85%). Technical failure was lower for oximetry than nasal pressure (5.8% vs 18.9% of tests).

Conclusions: In patients with high CV risk overnight single-channel oximetry and nasal pressure measurements may provide high diagnostic accuracy and offer an accessible alternative to full PSG.

Citing Articles

Sleep-related drugs utilization and suicide behaviors: a population-based study in China.

Zhang Y, Liu B, Li K, Jia H, Shi H, Liu Y BMC Public Health. 2025; 25(1):199.

PMID: 39833756 PMC: 11744836. DOI: 10.1186/s12889-025-21443-x.


Screening Tools for Obstructive Sleep Apnea in Pregnant Women: An Extended and Updated Systematic Review and Meta-analysis.

Amra B, Mansouri M, Soltaninejad F, Feizi A, Kaminska M Int J Prev Med. 2024; 15:31.

PMID: 39239303 PMC: 11376534. DOI: 10.4103/ijpvm.ijpvm_88_23.


Prevalence of Obstructive Sleep Apnea Using Home Sleep Test in Taiwan During the Coronavirus Disease Pandemic.

Chou T, Hsu H, Twu C, Huang W, Huang H, Weng S Nat Sci Sleep. 2023; 15:1107-1116.

PMID: 38149042 PMC: 10750777. DOI: 10.2147/NSS.S434278.


The changes of AHI after long-term CPAP in patients with comorbid OSA and cardiovascular disease.

Cheng Y, Ou Q, Chen B, Loffler K, McEvoy R, Xu Y Sleep Breath. 2022; 27(2):511-518.

PMID: 35562555 DOI: 10.1007/s11325-022-02633-y.


Associations Between Self-Reported Sleep Disturbances and Cognitive Impairment: A Population-Based Cross-Sectional Study.

Sun L, Li K, Zhang L, Zhang Y Nat Sci Sleep. 2022; 14:207-216.

PMID: 35210888 PMC: 8857964. DOI: 10.2147/NSS.S347658.