Evaluation of the Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Test and Identification of Rare Polymorphisms Potentially Affecting Assay Performance

Overview
Specialty Microbiology
Date 2010 Jun 25
PMID 20573864
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

We evaluated the FDA-approved Roche Cobas AmpliPrep/Cobas TaqMan (CAP/CTM) HIV-1 viral load assay for sensitivity, reproducibility, linearity, HIV-1 subtype detection, and correlation to the Roche Amplicor HIV-1 monitor test, version 1.5 (Amplicor). The limit of detection calculated by probit analysis was 23.8 copies/ml using the 2nd International WHO Standard and 30.8 copies/ml using Viral Quality Assurance (VQA) standard material. Serial dilutions of six patient samples were used to determine inter- and intra-assay reproducibility and linearity, which were very good (<8% coefficient of variation [CV]; between approximately 1.7 and 7.0 log(10) copies/ml). Subtype detection was evaluated in the CAP/CTM, Amplicor, and Bayer Versant HIV-1 bDNA 3.0 (Versant) assays using a commercially available panel. Versant averaged 0.829 log(10) copies/ml lower than CAP/CTM and Amplicor averaged 0.427 log(10) copies/ml lower than CAP/CTM for the subtype panel. Correlation with samples previously tested by Amplicor was excellent (R(2) = 0.884; average difference [Amplicor value minus CAP/CTM value], 0.008 log(10) copies/ml). Of the 305 HIV samples tested, 7 samples generated CAP/CTM titers between 1.0 and 2.75 log(10) copies/ml lower than those for Amplicor. Three of these samples revealed primer and probe mismatches that could account for the discrepancies. Otherwise, the CAP/CTM assay exhibits excellent sensitivity, dynamic range, reproducibility, and correlation with Amplicor in an automated format.

Citing Articles

The impact of dolutegravir-based combination antiretroviral therapy on the spermatozoa and fertility parameters of men living with human immunodeficiency virus.

Akang E, Dosumu O, Ogbenna A, Akpan U, Ezeukwu J, Odofin M Andrologia. 2022; 54(11):e14621.

PMID: 36261884 PMC: 9722517. DOI: 10.1111/and.14621.


Evaluation of a Trio Toscana Virus Real-Time RT-PCR Assay Targeting Three Genomic Regions within Nucleoprotein Gene.

Thirion L, Pezzi L, Pedrosa-Corral I, Sanbonmatsu-Gamez S, de Lamballerie X, Falchi A Pathogens. 2021; 10(3).

PMID: 33668339 PMC: 7996202. DOI: 10.3390/pathogens10030254.


Characterizing Patients with Very-Low-Level HIV Viremia: A Community-Based Study.

Helou E, Shenoi S, Kyriakides T, Landry M, Kozal M, Barakat L J Int Assoc Provid AIDS Care. 2016; 16(3):261-266.

PMID: 27903948 PMC: 5423832. DOI: 10.1177/2325957416680028.


The scanning electron microscope in microbiology and diagnosis of infectious disease.

Golding C, Lamboo L, Beniac D, Booth T Sci Rep. 2016; 6:26516.

PMID: 27212232 PMC: 4876401. DOI: 10.1038/srep26516.


Risk Minimization Measures for Blood Screening HIV-1 Nucleic Acid Amplification Technique Assays in Germany.

Chudy M, Kress J, Halbauer J, Heiden M, Funk M, Nubling C Transfus Med Hemother. 2014; 41(1):45-51.

PMID: 24659947 PMC: 3949608. DOI: 10.1159/000357103.


References
1.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

2.
Schumacher W, Frick E, Kauselmann M, Maier-Hoyle V, van der Vliet R, Babiel R . Fully automated quantification of human immunodeficiency virus (HIV) type 1 RNA in human plasma by the COBAS AmpliPrep/COBAS TaqMan system. J Clin Virol. 2007; 38(4):304-12. DOI: 10.1016/j.jcv.2006.12.022. View

3.
Holguin A, Lopez M, Molinero M, Soriano V . Performance of three commercial viral load assays, Versant human immunodeficiency virus type 1 (HIV-1) RNA bDNA v3.0, Cobas AmpliPrep/Cobas TaqMan HIV-1, and NucliSens HIV-1 EasyQ v1.2, testing HIV-1 non-B subtypes and recombinant variants. J Clin Microbiol. 2008; 46(9):2918-23. PMC: 2546710. DOI: 10.1128/JCM.02414-07. View

4.
Damond F, Roquebert B, Benard A, Collin G, Miceli M, Yeni P . Human immunodeficiency virus type 1 (HIV-1) plasma load discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 assays. J Clin Microbiol. 2007; 45(10):3436-8. PMC: 2045351. DOI: 10.1128/JCM.00973-07. View

5.
Oliver A, Pereira S, Clark D . Comparative evaluation of the automated Roche TaqMan real-time quantitative human immunodeficiency virus type 1 RNA PCR assay and the Roche AMPLICOR Version 1.5 conventional PCR assay. J Clin Microbiol. 2007; 45(11):3616-9. PMC: 2168503. DOI: 10.1128/JCM.00221-07. View