Low-dose, 128-slice, Dual-source CT Coronary Angiography: Accuracy and Radiation Dose of the High-pitch and the Step-and-shoot Mode

Overview
Journal Heart
Date 2010 Jun 12
PMID 20538669
Citations 66
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare the diagnostic accuracy and radiation doses of two low-dose protocols for coronary artery imaging with second-generation, dual-source CT in comparison with catheter angiography (CA).

Design, Setting And Patients: Prospective, single-centre study conducted in a referral centre enrolling 100 patients with low-to-intermediate risk and suspicion of coronary artery disease. All patients underwent contrast-enhanced, 128-slice, dual-source CT coronary angiography and CA. Patients were randomly assigned to two different low-dose CT protocols (each 100 kV/320 mA): in group A (n=50), CT was performed using the prospectively electrocardiography (ECG)-gated step-and-shoot (SAS) mode; in group B (n=50), CT was performed using the prospectively ECG-gated high-pitch mode (pitch 3.4). The image quality and presence or absence of significant coronary stenosis in all coronary segments were evaluated by two blinded and independent observers. CA served as the standard of reference.

Results: Sixty-one significant stenoses were found in group A, and 69 in group B. There was no significant difference in age (group A, 62+/-8 yrs; group B, 63+/-8 yrs; p=0.72), body mass index (group A, 26.4+/-3.1 kg/m2; group B, 25.9+/-2.8 kg/m2; p=0.41) and heart rate (HR) (group A, 58+/-8 bpm; group B, 56+/-10 bpm; p=0.66) between the groups. Diagnostic image quality was obtained in 98.6% (651/660) of segments in group A and in 98.9% (642/649) in group B, with no significant differences between groups. Sensitivity, specificity and positive and negative predictive values were 94%, 91%, 85% and 97% per-patient in group A, and 93%, 94%, 89% and 97% per-patient in group B (no significant differences). The effective radiation dose in group B (0.9+/-0.1 mSv) was significantly (p<0.01) lower than that in group A (1.4+/-0.4 mSv).

Conclusions: Both the high-pitch and the SAS mode for low-dose CT coronary angiography provide high accuracy for the assessment of significant coronary stenoses, while the high-pitch mode further significantly lowers the radiation dose.

Citing Articles

The effectiveness of coronary computed tomography angiography and functional testing for the diagnosis of obstructive coronary artery disease: results from the individual patient data Collaborative Meta-Analysis of Cardiac CT (COME-CCT).

Schlattmann P, Wieske V, Bressem K, Gotz T, Schuetz G, Andreini D Insights Imaging. 2024; 15(1):208.

PMID: 39143443 PMC: 11324632. DOI: 10.1186/s13244-024-01702-y.


Low-dose, high-pitch, spiral (FLASH) mode versus conventional sequential method for coronary artery calcium scoring: A derivation-validation study.

Pandey N, Chakraborty S, Verma M, Jagia P J Cardiovasc Thorac Res. 2024; 16(1):15-20.

PMID: 38584662 PMC: 10997979. DOI: 10.34172/jcvtr.31736.


A Novel Reconstruction Technique to Reduce Stair-Step Artifacts in Sequential Mode Coronary CT Angiography.

Moser L, Mergen V, Allmendinger T, Manka R, Eberhard M, Alkadhi H Invest Radiol. 2024; 59(9):622-628.

PMID: 38284879 PMC: 11827681. DOI: 10.1097/RLI.0000000000001066.


Computed tomography angiography versus Agatston score for diagnosis of coronary artery disease in patients with stable chest pain: individual patient data meta-analysis of the international COME-CCT Consortium.

Wieske V, Walther M, Dubourg B, Alkadhi H, Norgaard B, Meijs M Eur Radiol. 2022; 32(8):5233-5245.

PMID: 35267094 PMC: 9279219. DOI: 10.1007/s00330-022-08619-4.


In vivo radiation dosimetry and image quality of turbo-flash and retrospective dual-source CT coronary angiography.

Schicchi N, Mari A, Fogante M, Esposto Pirani P, Agliata G, Tosi N Radiol Med. 2019; 125(2):117-127.

PMID: 31686317 DOI: 10.1007/s11547-019-01103-y.