Coil Migration Through a Neuroform 3 Stent During Endovascular Coiling. A Case Report
Overview
Authors
Affiliations
A 43-year-old woman attended for stent assisted coiling. A Neuroform 30 x 4.5 mm stent had been successfully placed over the left periophthalmic aneurysm. During the coiling the first coil migrated through the crowns in the stent, lodging at the MCA bifurcation. We believe that the coil herniated through the overlying stent due to the carotid siphon curvature and the open cell design. Furthermore the distal markers of the stent impeded coil extraction with a MERCI device.
Maeda K, Motoie R, Karashima S, Otsuji R, Ren N, Nagaoka S Interv Neuroradiol. 2018; 24(6):643-649.
PMID: 29871560 PMC: 6259335. DOI: 10.1177/1591019918780465.
Extraction of a migrated coil from the Enterprise stent strut using a Solitaire AB stent.
Ge H, Jin H, Li Y, Lv X Neuroradiol J. 2016; 29(6):470-472.
PMID: 27558991 PMC: 5131764. DOI: 10.1177/1971400916666556.
Treatment of endovascular coil and stent migration using the merci retriever: report of three cases.
Kung D, Abel T, Madhavan K, Dalyai R, Dlouhy B, Liu W Case Rep Med. 2012; 2012:242101.
PMID: 22666264 PMC: 3362840. DOI: 10.1155/2012/242101.