Studies on Coagulation and the Development of an Automatic Computerized Bipolar Coagulator. Technical Note

Overview
Journal J Neurosurg
Specialty Neurosurgery
Date 1991 Jul 1
PMID 2045902
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

A new computerized bipolar coagulator is described in which tissue heating is switched off automatically when adequate vessel occlusion has been achieved, thus preventing overheating, undue tissue damage, cutting, and sticking of the forceps. Experiments with radiofrequency (rf) heating of albumin or arteries revealed an impedance minimum at the moment of coagulation. The attainment of this impedance minimum is transmitted electronically via a microprocessor to the coagulator, which automatically shuts off the rf energy supply. In experiments, adequate artery strength and avoidance of the drawbacks of conventional coagulation methods were achieved when rf heating was shut off soon after the impedance minimum was reached. Neither irrigation for cooling nor cleaning of the forceps tips was necessary. Electronic feedback through the same cables as used for coagulation enabled the use of conventional bipolar cables and forceps. The bipolar coagulator described can also be used for conventional bipolar coagulation under visual control. The microcomputer enables: 1) automatic coagulation cycles that start when tissue is picked up in the forceps and stop automatically on completion of the seal; 2) the change of power setting from a pedal and activation of automatic cycles by the pedal as described above or surgeon-controlled coagulation, which facilitates the use of alternative debridement with inactive forceps; 3) cable testing; and 4) negligible disturbance of the intraoperative monitoring equipment.

Citing Articles

Local thermal effect of power-on setting on monopolar coagulation: a three-dimensional electrothermal coupled finite element study.

Chen Y, Tu Y, Tsai Y, Tsai Y, Hsiao C Med Biol Eng Comput. 2022; 60(12):3525-3538.

PMID: 36229717 DOI: 10.1007/s11517-022-02689-8.


Bipolar vessel sealing: instrument contamination and wear have little effect on seal quality and success in a porcine in vitro model.

Wallwiener C, Junginger S, Zubke W, Brucker S, Enderle M, Neugebauer A Langenbecks Arch Surg. 2014; 399(7):863-71.

PMID: 25064363 DOI: 10.1007/s00423-014-1234-2.


Comparison of thermal coagulation profiles for bipolar forceps with different cooling mechanisms in a porcine model of spinal surgery.

Chen R, Than K, Wang A, Park P, Shih A Surg Neurol Int. 2013; 4:113.

PMID: 24083049 PMC: 3779387. DOI: 10.4103/2152-7806.117709.