Intraoperative Femur Fracture is Associated with Stem and Instrument Design in Primary Total Hip Arthroplasty

Overview
Publisher Wolters Kluwer
Specialty Orthopedics
Date 2010 Apr 14
PMID 20387021
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Intraoperative proximal femoral fracture is a risk in total hip arthroplasty (THA) with cementless stems with reported rates of 1.5-27.8%. Certain designs or designs associated with certain types of instrumentation may have higher rates.

Questions/purposes: We compared the incidence of proximal femoral fracture with two stem and instrument design systems.

Methods: We retrospectively reviewed 425 patients (457 hips) undergoing primary THAs using a supine, muscle-sparing approach between February 2007 and April 2009. In 439 cases, a monoblock, broach-only, tapered wedge design stem was used with a single low-profile cutting blade (cutting) calcar mill. Eighteen proximally modular, broach-only, tapered wedge design stems were used with a calcar mill with multiple deep cutting teeth (toothed).

Results: There were no proximal femoral fractures among the 439 hips using the cutting mill and two intraoperative proximal femur fractures in 18 hips using the toothed mill. Both fractures occurred during calcar milling when the cutting teeth engaged the bone, resulting in a torque of the broach.

Conclusions: Given the high intraoperative proximal femur fracture rate using the toothed design, and the absence of fractures using the cutting design, we caution against the use of the toothed style. The overall rate using a flat, wedge, tapered design (two of 457 hips or 0.4%) is lower than that associated with previously published designs. Fracture appears directly related to instrumentation.

Level Of Evidence: Level III, retrospective comparison study. See Guidelines for Authors for a complete description of Levels of Evidence.

Citing Articles

Periprosthetic Fractures of the Femur during the Operation of Primary Total Hip Arthroplasties.

Trentini C, Balsano M, Deeke M, Zaniolo F Rev Bras Ortop (Sao Paulo). 2024; 59(5):e745-e751.

PMID: 39649054 PMC: 11624928. DOI: 10.1055/s-0044-1779320.


What is the clinical utility of acoustic and vibrational analyses in uncemented total hip arthroplasty?.

Patel S, Hecht 2nd C, Homma Y, Kamath A Arthroplasty. 2024; 6(1):59.

PMID: 39623495 PMC: 11613849. DOI: 10.1186/s42836-024-00280-0.


Intraoperative Acetabular Fractures in Primary Total Hip Arthroplasty Management and Functional Outcomes.

Fontanellas-Fes A, Lizano-Diez X, Leon-Garcia A, Pardo I Pol A, Martinez-Diaz S, Andriola V Eur J Trauma Emerg Surg. 2024; 50(5):2615-2620.

PMID: 39251435 DOI: 10.1007/s00068-024-02636-2.


Analysis of Postoperative Complication and Revision Rates and Mid- to Long-Term Implant Survival in Primary Short-Stem Total Hip Arthroplasty.

Stauss R, Becker N, Savov P, Ettinger M, Seeber G J Clin Med. 2024; 13(13).

PMID: 38999347 PMC: 11242722. DOI: 10.3390/jcm13133779.


Periprosthetic fracture following arthroplasty for femoral neck fracture: is a cemented stem protective?.

Esper G, Meltzer-Bruhn A, Anil U, Schwarzkopf R, Macaulay W, Konda S Eur J Orthop Surg Traumatol. 2024; 34(4):1927-1935.

PMID: 38462554 DOI: 10.1007/s00590-023-03830-5.


References
1.
Schwartz Jr J, Mayer J, Engh C . Femoral fracture during non-cemented total hip arthroplasty. J Bone Joint Surg Am. 1989; 71(8):1135-42. View

2.
Kyle R, Crickard 3rd G . Periprosthetic fractures associated with total hip arthroplasty. Orthopedics. 1998; 21(9):982-4. DOI: 10.3928/0147-7447-19980901-15. View

3.
Graf R, Azizbaig-Mohajer M . Minimally invasive total hip replacement with the patient in the supine position and the contralateral leg elevated. Oper Orthop Traumatol. 2006; 18(4):317-29. DOI: 10.1007/s00064-006-1180-4. View

4.
Mallory T, Lombardi Jr A, Leith J, Fujita H, Hartman J, Capps S . Why a taper?. J Bone Joint Surg Am. 2002; 84-A Suppl 2:81-9. DOI: 10.2106/00004623-200200002-00010. View

5.
Mallory T, Lombardi Jr A, Leith J, Fujita H, Hartman J, Capps S . Minimal 10-year results of a tapered cementless femoral component in total hip arthroplasty. J Arthroplasty. 2001; 16(8 Suppl 1):49-54. DOI: 10.1054/arth.2001.28721. View