Background: Microvascular anastomosis is one of the more critical aspects of free flap surgery. A safe, effective, and expedient method for venous anastomosis minimizes flap ischemia time, is easier on the surgical team, and saves costly operating room time. The authors report on their experience using the Synovis microvascular anastomotic coupling device in 1000 consecutive venous anastomoses in free flap breast reconstruction.
Methods: The authors retrospectively reviewed 1000 consecutive venous anastomoses that were performed using the microvascular anastomotic coupler between July of 2002 and July of 2008. Data were obtained on flap type, recipient vessel, coupler size, incidence of venous thrombosis, timing of venous thrombosis, and morbidity as a result of venous thrombosis.
Results: All anastomoses were performed in an end-to-end fashion. There were 460 unilateral cases and 270 bilateral cases of breast reconstruction. Flap types included muscle-sparing free transverse rectus abdominis myocutaneous, deep inferior epigastric perforator, superficial inferior epigastric artery, superior gluteal artery perforator, and inferior gluteal artery perforator. The vast majority of the recipient vessels were the internal mammary or thoracodorsal vessels. Most of the couplers that were used were either 3 or 2.5 mm in diameter. Overall, there were six instances of venous thrombosis (rate of 0.6 percent). There were no total flap losses due to venous thrombosis in this series, although two patients had partial flap necrosis.
Conclusions: The patency rate for venous anastomoses performed with the microvascular coupler is excellent when compared with standard suture techniques and has the advantage of overall easier application.
Laminette P, Hardie R JFMS Open Rep. 2024; 10(2):20551169241265244.
PMID: 39184328 PMC: 11342331. DOI: 10.1177/20551169241265244.
Promny T, Huberth P, Muller-Seubert W, Promny D, Cai A, Horch R J Clin Med. 2024; 13(8).
PMID: 38673438 PMC: 11050223. DOI: 10.3390/jcm13082165.
Kumar Vishwakarma A, Sharma N, Kumar Dhiman N, Singh S, Jaiswara C, Swaraj S J Oral Biol Craniofac Res. 2024; 14(2):116-125.
PMID: 38313578 PMC: 10835319. DOI: 10.1016/j.jobcr.2024.01.005.
A simple free flap strategy using end-to-side anastomosis to the main vessels in injured extremity.
Motomiya M, Watanabe N, Ota M, Shimoda K, Kawamura D, Iwasaki N JPRAS Open. 2023; 38:48-59.
PMID: 37675275 PMC: 10477053. DOI: 10.1016/j.jpra.2023.08.003.
A Novel Vascular Anastomotic Coupling Device for End-to-End Anastomosis of Arteries and Veins.
Azapagic A, Agarwal J, Gale B, Li H, Nelson S, Shea J IEEE Trans Biomed Eng. 2023; 71(2):542-552.
PMID: 37639422 PMC: 10846801. DOI: 10.1109/TBME.2023.3308890.