Hepatic Vascular Flow Measurements by Phase Contrast MRI and Doppler Echography: a Comparative and Reproducibility Study

Overview
Date 2010 Feb 27
PMID 20187200
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To directly compare and study the variability of parameters related to hepatic blood flow measurements using 3 T phase-contrast magnetic resonance imaging (PC-MRI) and Doppler ultrasound (US).

Materials And Methods: Nine healthy subjects were studied. Blood velocities and flow rate measurements were performed in the portal vein and the proper hepatic artery. MR studies were performed using a 3 T imager. Gradient-echo fast phase contrast sequences were used with both cardiac and respiratory gating. MR and Doppler flow parameters were extracted and compared. Two methods of calculation were used for Doppler flow rate analysis.

Results: Compared to Doppler US, PC-MRI largely underestimated hepatic flow data with lower variability and higher reproducibility. This reproducibility was more pronounced in the portal vein than in the proper hepatic artery associated with poorer velocity correlations. Total hepatic flow values were 1239 +/- 223 mL/min and 1595 +/- 521 mL/min for PC-MRI and Doppler US, respectively.

Conclusion: Free-breathing PC-MRI can provide reliable noninvasive measurement of hepatic flow parameters compared to Doppler US. The MR technique could help to improve Doppler flow calculations, thereby allowing standardization of protocols, particularly for applications in disease.

Citing Articles

Reference values for 4D flow magnetic resonance imaging of the portal venous system.

Huang A, Roberts G, Roldan-Alzate A, Wieben O, Reeder S, Oechtering T Abdom Radiol (NY). 2023; 48(6):2049-2059.

PMID: 37016247 PMC: 10518803. DOI: 10.1007/s00261-023-03892-3.


4D Flow MRI in the portal venous system: imaging and analysis methods, and clinical applications.

Hyodo R, Takehara Y, Naganawa S Radiol Med. 2022; 127(11):1181-1198.

PMID: 36123520 PMC: 9587937. DOI: 10.1007/s11547-022-01553-x.


Portal Vein Pulsatility Index as a Potential Risk of Venous Congestion Assessed by Magnetic Resonance Imaging: A Prospective Study on Healthy Volunteers.

Abou-Arab O, Beyls C, Moussa M, Huette P, Beaudelot E, Guilbart M Front Physiol. 2022; 13:811286.

PMID: 35574483 PMC: 9101294. DOI: 10.3389/fphys.2022.811286.


Network Meta-Analysis: Noninvasive Imaging Modalities for Identifying Clinically Significant Portal Hypertension.

Hai Y, Chong W, Eisenbrey J, Forsberg F Dig Dis Sci. 2021; 67(7):3313-3326.

PMID: 34275089 PMC: 8761784. DOI: 10.1007/s10620-021-07168-y.


In vitro evaluation of cerebrospinal fluid velocity measurement in type I Chiari malformation: repeatability, reproducibility, and agreement using 2D phase contrast and 4D flow MRI.

Williams G, Thyagaraj S, Fu A, Oshinski J, Giese D, Bunck A Fluids Barriers CNS. 2021; 18(1):12.

PMID: 33736664 PMC: 7977612. DOI: 10.1186/s12987-021-00246-3.