» Articles » PMID: 20173650

Reproducibility of Retinal Nerve Fiber Thickness Measurements Using the Test-retest Function of Spectral OCT/SLO in Normal and Glaucomatous Eyes

Overview
Journal J Glaucoma
Date 2010 Feb 23
PMID 20173650
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the reproducibility of retinal nerve fiber layer (RNFL) thickness measurements in normal and glaucoma patients using the spectral optical coherence tomography (OCT)/scanning laser ophthalmosope (SLO).

Methods: One experienced operator measured peripapillary RNFL thickness 3 times in succession in 98 normal participants and 79 glaucoma patients by using the test-retest function of the spectral OCT/SLO. The intraclass correlation coefficient (ICC), coefficient of variation, and test-retest variability (TRV) were analyzed for the mean RNFL thickness, and the RNFL thickness in each quadrant and at each clock-hour, in randomly selected eyes. The correlation coefficient was calculated between the TRV and the mean RNFL thickness, and between the TRV and the mean deviation of the Humphrey visual field.

Results: The ICC for mean RNFL thickness was 0.988 in the normal group and 0.993 in the glaucoma group. The ICC for RNFL thickness in all quadrants and clock-hours in both normal and glaucoma patients ranged from 0.951 to 0.992. The coefficient of variation was between 1.9% and 5.0% in the normal group and between 2.0% and 5.3% in the glaucoma group. The TRV was between 4.3 and 11.5 μm in the normal group and between 3.8 and 10.0 μm in the glaucoma group. The TRV showed no correlation with the mean RNFL thickness or the mean deviation (Pearson r>0.15, P>0.2).

Conclusions: Spectral OCT/SLO equipped with a test-retest function showed excellent reproducibility with regard to RNFL thickness measurements. The severity of glaucoma did not affect the reproducibility. These findings suggest that spectral OCT/SLO fulfills one of the basic premises for the diagnosis and progression analysis of glaucoma.

Citing Articles

Biological and Methodological Variability in Retinal Nerve Fiber Layer OCT: The Framingham Heart Study.

Almidani L, Sabharwal J, Shahidzadeh A, Collazo Martinez A, Ting S, Vaidya B Ophthalmol Sci. 2024; 4(6):100549.

PMID: 39161752 PMC: 11331915. DOI: 10.1016/j.xops.2024.100549.


Evaluation of Retinal Nerve Fiber Layer and Macular Ganglion Cell Layer Thickness in Relation to Optic Disc Size.

Storp J, Storp N, Danzer M, Eter N, Biermann J J Clin Med. 2023; 12(7).

PMID: 37048556 PMC: 10095471. DOI: 10.3390/jcm12072471.


Reproducibility of Neuroretinal Rim Measurements Obtained from High-Density Spectral Domain Optical Coherence Tomography Volume Scans.

Kim J, Men C, Ratanawongphaibul K, Papadogeorgou G, Tsikata E, Ben-David G Clin Ophthalmol. 2022; 16:2595-2608.

PMID: 35992568 PMC: 9384967. DOI: 10.2147/OPTH.S369807.


Aragonite toothpaste for management of dental calculus: A double-blinded randomized controlled clinical trial.

Al-Hashedi A, Dubreuil N, Schwinghamer T, Dorzhiyeva S, Anweigi L, Emami E Clin Exp Dent Res. 2022; 8(4):863-874.

PMID: 35419985 PMC: 9382037. DOI: 10.1002/cre2.559.


Disc Hemorrhages Are Associated With Localized Three-Dimensional Neuroretinal Rim Thickness Progression in Open-Angle Glaucoma.

Margeta M, Ratanawongphaibul K, Tsikata E, Zemplenyi M, Ondeck C, Kim J Am J Ophthalmol. 2021; 234:188-198.

PMID: 34214455 PMC: 8714862. DOI: 10.1016/j.ajo.2021.06.021.