Hemoglobin A1c Point-of-care Assays; a New World with a Lot of Consequences!
Overview
Affiliations
Background: Point-of-care instruments for the measurement of hemoglobin A1c (HbA1c) may improve the glycemic control of people with diabetes by providing a rapid result if the performance of the instruments used is acceptable. A 0.5% HbA1c difference between successive results is considered a clinically relevant change. With this in mind, the In2it from Bio-Rad and the DCA Vantage from Siemens were evaluated according to Clinical and Laboratory Standards Institute (CLSI) protocols.
Methods: The CLSI protocols EP-5 and EP-9 were applied to investigate precision, accuracy, and bias. The bias was compared with three certified secondary reference measurement procedures. Differences between capillary and venous blood were investigated by an end-user group consisting of nurse practitioners at a diabetes care center.
Results: At HbA1c levels of 5.1 and 11.2%, total coefficients of variation (CV) for the In2it were 4.9 and 3.3%, respectively, and for the DCA Vantage were 1.7 to 1.8% and 3.7 to 5.5% depending on the lot number of the cartridges. Method comparisons showed significant lot number-dependent results for the In2it and the DCA Vantage compared with the three reference methods. No overall difference was observed between capillary and venous blood for both methods.
Conclusion: Performance results of the In2it and the DCA Vantage showed variable and lot number-dependent results. To maintain the interlaboratory CV of 5% for HbA1c, the Clinical Laboratory Improvement Amendments rules for waived point-of-care instruments should be revised. An obligation for participating in external quality schemes and taking adequate action should be considered for POC instruments that perform poorly.
Point-of-Care HbA1c in Clinical Practice: Caveats and Considerations for Optimal Use.
Sacks D, Kirkman M, Little R Diabetes Care. 2024; 47(7):1104-1110.
PMID: 38552140 PMC: 11208753. DOI: 10.2337/dci23-0040.
Ikem R, Enikuomehin A, Soyoye D, Kolawole B Pan Afr Med J. 2023; 43:148.
PMID: 36785682 PMC: 9922079. DOI: 10.11604/pamj.2022.43.148.36586.
Liddy A, Grundy S, Sreenan S, Tormey W Ir J Med Sci. 2022; 192(1):169-176.
PMID: 35362846 PMC: 9892076. DOI: 10.1007/s11845-022-02967-2.
Enikuomehin A, Kolawole B, Soyoye O, Adebayo J, Ikem R Afr Health Sci. 2021; 20(1):294-307.
PMID: 33402918 PMC: 7750067. DOI: 10.4314/ahs.v20i1.35.
Methods, units and quality requirements for the analysis of haemoglobin A1c in diabetes mellitus.
Penttila I, Penttila K, Holm P, Laitinen H, Ranta P, Torronen J World J Methodol. 2016; 6(2):133-42.
PMID: 27376018 PMC: 4921944. DOI: 10.5662/wjm.v6.i2.133.