» Articles » PMID: 20138082

A Novel Screening Method for Influenza Patients Using a Newly Developed Non-contact Screening System

Overview
Journal J Infect
Date 2010 Feb 9
PMID 20138082
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: In places of mass gathering, rapid infection screening prior to definite diagnosis is vital during the epidemic season of a novel influenza. In order to assess the possibility of clinical application of a newly developed non-contact infection screening system, we conducted screening for influenza patients.

Materials And Methods: The system is operated by a screening program via a linear discriminant analysis using non-contact derived variables, i.e., palmar pulse derived from a laser Doppler blood-flow meter, respiration rate determined by a 10-GHz microwave radar, and average facial temperature measured by thermography. The system was tested on 57 seasonal influenza (2008-2009) patients (35.7 degrees C < or = body temperature < or = 38.3 degrees C, 19-40 years) and 35 normal control subjects (35.5 degrees C < or = body temperature < or = 36.9 degrees C, 21-35 years) at the Japan Self-defense Forces Central Hospital.

Results: A significant linear discriminant function (p < 0.001) was determined to distinguish the influenza group from the control group (Mahalanobis D-square = 6.5, classification error rate > 10%). The system had a positive predictive value (PPV) of 93%, which is higher than the PPV value (PPV < or = 65.4%) reported in the recent summary of studies using only thermography performed mainly in hospitals.

Conclusions: The proposed system appears promising for application in accurate screening for influenza patients at places of mass gathering.

Citing Articles

Enhancing Salty Taste Perception in Stroke Patients via Anodal Electrical Stimulation to the Chin.

Katsuki M, Fukushima T, Wada N, Goto T, Imai A, Hanaoka Y Foods. 2025; 13(24.

PMID: 39767033 PMC: 11675263. DOI: 10.3390/foods13244087.


Systematic Literature Review Regarding Heart Rate and Respiratory Rate Measurement by Means of Radar Technology.

Liebetruth M, Kehe K, Steinritz D, Sammito S Sensors (Basel). 2024; 24(3.

PMID: 38339721 PMC: 10857015. DOI: 10.3390/s24031003.


Use of Infrared Thermography in Medical Diagnosis, Screening, and Disease Monitoring: A Scoping Review.

Kesztyus D, Brucher S, Wilson C, Kesztyus T Medicina (Kaunas). 2023; 59(12).

PMID: 38138242 PMC: 10744680. DOI: 10.3390/medicina59122139.


Non-Contact Infrared Thermometers and Thermal Scanners for Human Body Temperature Monitoring: A Systematic Review.

Zhao Y, Bergmann J Sensors (Basel). 2023; 23(17).

PMID: 37687902 PMC: 10490756. DOI: 10.3390/s23177439.


Statistical Study on Human Temperature Measurement by Infrared Thermography.

Svantner M, Lang V, Skala J, Kohlschutter T, Honner M, Muzika L Sensors (Basel). 2022; 22(21).

PMID: 36366093 PMC: 9654967. DOI: 10.3390/s22218395.


References
1.
Lipsitch M, Cohen T, Cooper B, Robins J, Ma S, James L . Transmission dynamics and control of severe acute respiratory syndrome. Science. 2003; 300(5627):1966-70. PMC: 2760158. DOI: 10.1126/science.1086616. View

2.
Bitar D, Goubar A, Desenclos J . International travels and fever screening during epidemics: a literature review on the effectiveness and potential use of non-contact infrared thermometers. Euro Surveill. 2009; 14(6). View

3.
Gottfredsson M, Halldorsson B, Jonsson S, Kristjansson M, Kristjansson K, Kristinsson K . Lessons from the past: familial aggregation analysis of fatal pandemic influenza (Spanish flu) in Iceland in 1918. Proc Natl Acad Sci U S A. 2008; 105(4):1303-8. PMC: 2234133. DOI: 10.1073/pnas.0707659105. View

4.
Kikuchi M, Ishihara M, Matsui T, Wakisaka H, Ashida H, Sato S . Biomedical engineering's contribution to defending the homeland. IEEE Eng Med Biol Mag. 2004; 23(1):175-86. DOI: 10.1109/memb.2004.1297190. View

5.
Leclerc J, Pu Q, Corseaux D, Haddad E, Decoene C, Bordet R . A single endotoxin injection in the rabbit causes prolonged blood vessel dysfunction and a procoagulant state. Crit Care Med. 2000; 28(11):3672-8. DOI: 10.1097/00003246-200011000-00023. View