Metal-backed Acetabular Components with Conventional Polyethylene: a Review of 9113 Primary Components with a Follow-up of 20 Years

Overview
Date 2010 Feb 5
PMID 20130308
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

The Norwegian Arthroplasty Register has shown that several designs of uncemented femoral stems give good or excellent survivorship. The overall findings for uncemented total hip replacement however, have been disappointing because of poor results with the use of metal-backed acetabular components. In this study, we exclusively investigated the medium-to long-term performance of primary uncemented metal-backed acetabular components. A total of 9113 primary uncemented acetabular components were implanted in 7937 patients between 1987 and 2007. These were included in a prospective, population-based observational study. All the implants were modular and metal-backed with ultra-high-molecular-weight polyethylene liners. The femoral heads were made of stainless steel, cobalt-chrome (CoCr) alloy or alumina ceramic. In all, seven different designs of acetabular component were evaluated by the Kaplan-Meier survivorship method and Cox regression analysis. Most acetabular components performed well up to seven years. When the endpoint was revision of the acetabular component because of aseptic loosening, the survival ranged between 87% and 100% at ten years. However, when the endpoint was revision for any reason, the survival estimates were 81% to 92% for the same implants at ten years. Aseptic loosening, wear, osteolysis and dislocation were the main reasons for the relatively poor overall performance of the acetabular components. Prostheses with alumina heads performed slightly better than those with stainless steel or CoCr alloy in subgroups. Whereas most acetabular components performed well at seven years, the survivorship declined with longer follow-up. Fixation was generally good. None of the metal-backed uncemented acetabular components with ultra-high-molecular-weight polyethylene liners in our study had satisfactory long-term results because of high rates of wear, osteolysis, aseptic loosening and dislocation.

Citing Articles

Challenges and Long-Term Outcomes of Cementless Total Hip Arthroplasty in Patients Under 30: A 24-Year Follow-Up Study with a Minimum 8-Year Follow-Up, Focused on Developmental Dysplasia of the Hip.

Drobniewski M, Gonera B, Olewnik L, Borowski A, Ruzik K, Triantafyllou G J Clin Med. 2024; 13(21).

PMID: 39518730 PMC: 11546925. DOI: 10.3390/jcm13216591.


Early unrestricted vs. partial weight bearing after uncemented total hip arthroplasty: a systematic review and meta-analysis.

Huang L, Han W, Qi W, Zhang X, Lv Z, Lu Y Front Surg. 2023; 10:1225649.

PMID: 38033533 PMC: 10684916. DOI: 10.3389/fsurg.2023.1225649.


Long-term follow-up and survival of delayed total hip arthroplasty following acetabular fracture: a matched cohort study of 552 cases from the Norwegian Arthroplasty Register.

Kirkeboe R, Nordsletten L, Madsen J, Dybvik E, Lie S, Hallan G Hip Int. 2023; 34(3):396-401.

PMID: 37990005 PMC: 11071598. DOI: 10.1177/11207000231212884.


Fixation, sex, and age: highest risk of revision for uncemented stems in elderly women - data from 66,995 primary total hip arthroplasties in the Norwegian Arthroplasty Register.

Dale H, Borsheim S, Kristensen T, Fenstad A, Gjertsen J, Hallan G Acta Orthop. 2019; 91(1):33-41.

PMID: 31663413 PMC: 7006785. DOI: 10.1080/17453674.2019.1682851.


Impact of hip arthroplasty registers on orthopaedic practice and perspectives for the future.

Varnum C, Pedersen A, Rolfson O, Rogmark C, Furnes O, Hallan G EFORT Open Rev. 2019; 4(6):368-376.

PMID: 31210974 PMC: 6549115. DOI: 10.1302/2058-5241.4.180091.