In Vitro Biomechanical Analysis of a New Lumbar Low-profile Locking Screw-plate Construct Versus a Standard Top-loading Cantilevered Pedicle Screw-rod Construct: Technical Report

Overview
Journal Neurosurgery
Specialty Neurosurgery
Date 2010 Jan 21
PMID 20087110
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: A standard top-loading lumbar pedicle screw-rod system is compared with a pedicle screw-plate system with smaller-diameter screws, more medial entry, and lower profile to assess the relative stability, strength, and resistance to fatigue of the 2 systems.

Methods: Seven human cadaveric specimens were studied with each surgical construct. Nondestructive, nonconstraining pure moments were applied to specimens to induce flexion, extension, lateral bending, and axial rotation while recording L5-S1 motion optoelectronically. After initial tests, specimens were fatigued for 10,000 cycles and retested to assess early postoperative loosening. Specimens were then loaded to failure in hyperextension.

Results: The standard screw-rod construct reduced range of motion to a mean of 20% of normal, whereas the screw-plate construct reduced range of motion to 13% of normal. Differences between systems were not significant in any loading mode (P > 0.06). The 14% loosening of the screw-rod system with fatigue was not significantly different from the 10% loosening observed with the screw-plate system (P > 0.15). Mean failure loads of 30 Nm for screw-rod and 37 Nm for screw-plate were also not significantly different (P = 0.38).

Conclusion: Posterior fixation at L5-S1 using the low-profile screw-plate system offers stability, resistance to fatigue, and resistance to failure equivalent to fixation using a standard cantilevered pedicle screw-rod system.

Citing Articles

A spontaneous anterior fusion of lumbar spine after posterolateral lumbar fusion with pedicle screw-plate system.

Kraiwattanapong C, Chandrsawang I, Keorochana G, Rajinda P, Chanplakorn P, Leelapattana P BMC Musculoskelet Disord. 2025; 26(1):95.

PMID: 39893374 PMC: 11786541. DOI: 10.1186/s12891-024-08184-5.


Finite Element Analysis of a New Pedicle Screw-Plate System for Minimally Invasive Transforaminal Lumbar Interbody Fusion.

Li J, Shang J, Zhou Y, Li C, Liu H PLoS One. 2015; 10(12):e0144637.

PMID: 26649749 PMC: 4674154. DOI: 10.1371/journal.pone.0144637.


A novel blasted and grooved low profile pedicle screw able to resist high compression bending loads.

Kuh S, Kim Y, Choi H, Kim K, Park J, Jeong H Korean J Spine. 2015; 9(2):61-5.

PMID: 25983790 PMC: 4432362. DOI: 10.14245/kjs.2012.9.2.61.


Novel pedicle screw and plate system provides superior stability in unilateral fixation for minimally invasive transforaminal lumbar interbody fusion: an in vitro biomechanical study.

Li J, Xiao H, Zhu Q, Zhou Y, Li C, Liu H PLoS One. 2015; 10(3):e0123134.

PMID: 25807513 PMC: 4373727. DOI: 10.1371/journal.pone.0123134.


The current testing protocols for biomechanical evaluation of lumbar spinal implants in laboratory setting: a review of the literature.

Gonzalez-Blohm S, Doulgeris J, Lee 3rd W, Shea T, Aghayev K, Vrionis F Biomed Res Int. 2015; 2015:506181.

PMID: 25785272 PMC: 4345069. DOI: 10.1155/2015/506181.