Use of a Novel Needle Wire in Patients Undergoing Transseptal Puncture Associated with Severe Septal Tenting

Overview
Publisher Springer
Date 2010 Jan 15
PMID 20072804
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: An increasing number of patients undergo left atrial ablation procedures, since several approaches have proven efficacy in the treatment of atrial fibrillation. Although transseptal catheterization was generally shown be a safe technique, it harbors the principal risk of cardiac injury. Therefore, there is a need for a safe and effective tool to enable transseptal puncture in difficult cases as well.

Methods And Results: In 158 consecutive patients, a transseptal puncture was intended for mapping and ablation of left atrial tachycardias. In seven patients of this series transseptal puncture using different sheaths and needle designs, the operators failed to cross the interatrial septum as a result of severe tenting. Three patients were known to have a septal aneurysm; a redo procedure was performed in two patients. In the remaining patients, there was no obvious explanation for the difficulty in crossing the interatrial septum conventionally. In all seven patients, a 120-cm-long nitinol guidewire ("needle wire") with a 0.014-inch diameter was used to cross the atrial septum with the following idea: after tenting the fossa ovalis with the transseptal dilator and the Brockenbrough needle positioned just inside the tip, effortless advancement of the needle wire perforates the membranous fossa. Unsupported by the needle and dilator, the tip of the wire immediately assumes a "J" shape, rendering it incapable of further tissue penetration after its entry into the left atrium. In all seven patients, the needle wire could be placed into a left pulmonary vein. In five patients, a single attempt was sufficient to reach the left atrium with the wire, two patients needed two and three attempts, respectively. No complications occurred.

Conclusions: Additional use of a needle wire to perform transseptal puncture in a subset of patients at higher risk for complications appears safe and effective.

Citing Articles

Complication prevention in ablation procedures: How to perform transseptal puncture safely in case of atrial septum aneurysm.

Stockigt F, Eberhardt F, Horlitz M HeartRhythm Case Rep. 2020; 5(11):529-533.

PMID: 31890566 PMC: 6926205. DOI: 10.1016/j.hrcr.2019.06.004.


Modified Transseptal Puncture Technique in Challenging Septa: A Randomized Comparison to Conventional Technique.

Kataria V, Berte B, Vandekerckhove Y, Tavernier R, Duytschaever M Biomed Res Int. 2017; 2017:2351925.

PMID: 28251152 PMC: 5303838. DOI: 10.1155/2017/2351925.


Utility and safety of the SafeSept™ transseptal guidewire for electrophysiology studies with catheter ablation in pediatric and congenital heart disease.

Knadler J, Anderson J, Chaouki A, Czosek R, Connor C, Knilans T J Interv Card Electrophysiol. 2017; 48(3):369-374.

PMID: 28091832 DOI: 10.1007/s10840-017-0224-z.

References
1.
Tomlinson D, Sabharwal N, Bashir Y, Betts T . Interatrial septum thickness and difficulty with transseptal puncture during redo catheter ablation of atrial fibrillation. Pacing Clin Electrophysiol. 2008; 31(12):1606-11. DOI: 10.1111/j.1540-8159.2008.01233.x. View

2.
Saliba W, Thomas J . Intracardiac echocardiography during catheter ablation of atrial fibrillation. Europace. 2008; 10 Suppl 3:iii42-7. DOI: 10.1093/europace/eun233. View

3.
Sakata Y, Feldman T . Transcatheter creation of atrial septal perforation using a radiofrequency transseptal system: novel approach as an alternative to transseptal needle puncture. Catheter Cardiovasc Interv. 2005; 64(3):327-32. DOI: 10.1002/ccd.20284. View

4.
Fagundes R, Mantica M, De Luca L, Forleo G, Pappalardo A, Avella A . Safety of single transseptal puncture for ablation of atrial fibrillation: retrospective study from a large cohort of patients. J Cardiovasc Electrophysiol. 2007; 18(12):1277-81. DOI: 10.1111/j.1540-8167.2007.00958.x. View

5.
Hu Y, Tai C, Lin Y, Chang S, Lo L, Wongcharoen W . The change in the fluoroscopy-guided transseptal puncture site and difficult punctures in catheter ablation of recurrent atrial fibrillation. Europace. 2008; 10(3):276-9. DOI: 10.1093/europace/eun013. View