Measurement of Fetal Head Descent Using the 'angle of Progression' on Transperineal Ultrasound Imaging is Reliable Regardless of Fetal Head Station or Ultrasound Expertise

Overview
Date 2010 Jan 14
PMID 20069668
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To assess whether ultrasound experience or fetal head station affects the reliability of measurement of fetal head descent using the angle of progression on intrapartum ultrasound images obtained by a single experienced operator, and to determine reliability of measurements when images were acquired by different operators with variable ultrasound experience.

Methods: One experienced obstetrician performed 44 transperineal ultrasound examinations of women at term and in prolonged second stage of labor with the fetus in the occipitoanterior position. Three midwives without ultrasound experience, three obstetricians with < 5 years' experience and three obstetricians with > 10 years' experience measured fetal head descent based on the angle of progression in the images obtained. The angle of progression was measured by two obstetricians in independent ultrasound examinations of 24 laboring women at term with the fetus in the cephalic position to allow assessment of the reliability of image acquisition. Intraclass correlation coefficients (ICCs) with 95% confidence interval (CI) were used to evaluate interobserver reliability and Bland-Altman analysis was used to assess interobserver agreement.

Results: In total, 444 measurements were performed and compared. Interobserver reliability with respect to offline image analysis was substantial (overall ICC, 0.72; 95% CI, 0.63-0.81). ICCs were 0.82 (95% CI, 0.70-0.89), 0.81 (95% CI, 0.71-0.88) and 0.61 (95% CI, 0.43-074) for observers with > 10 years', < 5 years' and no ultrasound experience, respectively. There were no significant differences between ICCs among observer groups according to ultrasound experience. Fetal head station did not affect reliability. Bland-Altman analysis indicated reasonable agreement between measurements obtained by two different operators with > 10 years' and < 5 years' ultrasound experience (bias, -1.09 degrees ; 95% limits of agreement, -8.76 to 6.58). The reliability of measurement of the angle of progression following separate image acquisition by two experienced operators was similar to the reliability of offline image analysis (ICC, 0.86; 95% CI, 0.70-0.93).

Conclusions: Measurement of the angle of progression on transperineal ultrasound imaging is reliable regardless of fetal head station or the clinician's level of ultrasound experience.

Citing Articles

Measurement of angle of progression by trans perineal-ultrasound in labour to predict mode of delivery in nulliparous women.

Masood H, Abid R, Ahmed M, Mehmood M Pak J Med Sci. 2025; 41(1):147-150.

PMID: 39867802 PMC: 11755286. DOI: 10.12669/pjms.41.1.9930.


A framework for computing angle of progression from transperineal ultrasound images for evaluating fetal head descent using a novel double branch network.

Bai J, Sun Z, Yu S, Lu Y, Long S, Wang H Front Physiol. 2022; 13:940150.

PMID: 36531181 PMC: 9755498. DOI: 10.3389/fphys.2022.940150.


Multitask Deep Neural Network for the Fully Automatic Measurement of the Angle of Progression.

Lu Y, Zhi D, Zhou M, Lai F, Chen G, Ou Z Comput Math Methods Med. 2022; 2022:5192338.

PMID: 36092792 PMC: 9462992. DOI: 10.1155/2022/5192338.


Intrapartum ultrasound use in clinical practice as a predictor of delivery mode during prolonged second stage of labor.

Katzir T, Brezinov Y, Khairish E, Hadad S, Vaisbuch E, Levy R Arch Gynecol Obstet. 2022; 307(3):763-770.

PMID: 35576076 DOI: 10.1007/s00404-022-06469-5.


Effects of Sex, Age and Height on Symphysis-Ischial Spine Distance Measured on a Pelvic CT.

Sanchez Garcia D, Aguado Del Hoyo A, Sanchez Perez M, Larroca S, Ruiz Martin Y, Gordillo Gutierrez I J Clin Med. 2022; 11(9).

PMID: 35566521 PMC: 9104774. DOI: 10.3390/jcm11092395.