Accuracy of Bioelectrical Impedance Consumer Devices for Measurement of Body Composition in Comparison to Whole Body Magnetic Resonance Imaging and Dual X-ray Absorptiometry

Overview
Journal Obes Facts
Publisher Karger
Specialty Endocrinology
Date 2010 Jan 8
PMID 20054195
Citations 89
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare body composition determined by bioelectrical impedance (BIA) consumer devices against criterion estimates determined by whole body magnetic resonance imaging (MRI) and dual energy X-ray absorptiometry (DXA) in healthy normal weight, overweight and obese adults.

Methods: In 106 adults (54 females, 52 males, age 54.2 +/- 16.1 years, BMI 25.8 +/- 4.4 kg/m(2)) fat mass (FM), skeletal muscle mass (SM), total body bone-free lean mass (TBBLM), and level of visceral fat mass (VF) were estimated by 3 single-frequency bipedal (foot-to-foot) and one tretrapolar BIA device, and compared to body composition measured by MRI and DXA. Bland-Altman and simple linear regression analyses were used to determine agreement between methods.

Results: %FMDXA, SMMRI or TBBLMDXA showed good relative and absolute agreement with two bipolar and one tetrapolar instrument (r(2) = 0.92-0.96; all p < 0.001; mean bias <1.5 %FM and <1 kg SM or TBBLM) and less relative and absolute agreement for another bipolar device (r(2) = 0.82 and 0.84, mean bias approximately 3 %FM and approximately 3 kg SM). The 95% limits of agreement (bias +/- 2 SD) were narrowest for the tetrapolar device (-6.59 to 4.61 %FM and -4.62 to 4.74 kg SM) and widest for bipolar instruments (up to -14.54 to 8.58 %FM and -9.52 to 3.92 kg SM). Systematic biases for %FM were found for all bipedal devices, but not for the tetrapolar instrument.

Conclusion: Because of the lower agreement between foot-to-foot BIA and DXA or MRI for the assessment of body composition in individuals, tetrapolar electrode arrangement should be preferred for individual or public use. Bipolar devices provide accurate results for field studies with group estimation.

Citing Articles

Consumption of lettuce with seaweed extract biostimulant application improved iron homeostasis in a randomized interventional trial of healthy individuals.

Di Gaudio F, Vasto S, Sabatino L, Ferrantelli V, Macaluso A, Caldarella R Sci Rep. 2025; 15(1):7799.

PMID: 40050333 PMC: 11885599. DOI: 10.1038/s41598-025-91380-7.


Unveiling the predictive role of motor competence and physical fitness on inhibitory control in preschool children: a cross-sectional study.

Canli U, Gut A, Sevinc S, Deveci M, Sendil A, Yaman G BMC Pediatr. 2025; 25(1):153.

PMID: 40022081 PMC: 11869643. DOI: 10.1186/s12887-025-05475-1.


Chemosensory and cardiometabolic improvements after a fasting-mimicking diet: A randomized cross-over clinical trial.

Micarelli A, Mrakic-Sposta S, Vezzoli A, Malacrida S, Caputo S, Micarelli B Cell Rep Med. 2025; 6(2):101971.

PMID: 39970875 PMC: 11866515. DOI: 10.1016/j.xcrm.2025.101971.


Acute effects of virtual reality exercise bike games on psychophysiological outcomes in college North-African adolescents with cerebral palsy: A randomized clinical trial.

Soudani M, Farhat F, Ghram A, Ben Saad H, Chlif M F1000Res. 2025; 12():1597.

PMID: 39834874 PMC: 11745216. DOI: 10.12688/f1000research.143189.2.


Combination of body mass index and body fat percentage in middle and late pregnancy to predict pregnancy outcomes in patients with gestational diabetes in Wenzhou, China: a single-centre retrospective cohort study.

Chen B, Chen L, Zhao X, You T, Zheng Z, Chen Y BMJ Open. 2024; 14(11):e086703.

PMID: 39521470 PMC: 11551990. DOI: 10.1136/bmjopen-2024-086703.


References
1.
Pateyjohns I, Brinkworth G, Buckley J, Noakes M, Clifton P . Comparison of three bioelectrical impedance methods with DXA in overweight and obese men. Obesity (Silver Spring). 2006; 14(11):2064-70. DOI: 10.1038/oby.2006.241. View

2.
Jebb S, Siervo M, Murgatroyd P, Evans S, Fruhbeck G, Prentice A . Validity of the leg-to-leg bioimpedance to estimate changes in body fat during weight loss and regain in overweight women: a comparison with multi-compartment models. Int J Obes (Lond). 2006; 31(5):756-62. DOI: 10.1038/sj.ijo.0803475. View

3.
Piers L, Soares M, Frandsen S, ODea K . Indirect estimates of body composition are useful for groups but unreliable in individuals. Int J Obes Relat Metab Disord. 2000; 24(9):1145-52. DOI: 10.1038/sj.ijo.0801387. View

4.
Xie X, Kolthoff N, Barenholt O, Nielsen S . Validation of a leg-to-leg bioimpedance analysis system in assessing body composition in postmenopausal women. Int J Obes Relat Metab Disord. 1999; 23(10):1079-84. DOI: 10.1038/sj.ijo.0801034. View

5.
Segal K, Van Loan M, Fitzgerald P, Hodgdon J, Van Itallie T . Lean body mass estimation by bioelectrical impedance analysis: a four-site cross-validation study. Am J Clin Nutr. 1988; 47(1):7-14. DOI: 10.1093/ajcn/47.1.7. View