A Novel Method for Patient Exit and Entrance Dose Prediction Based on Water Equivalent Path Length Measured with an Amorphous Silicon Electronic Portal Imaging Device

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2009 Dec 19
PMID 20019398
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

In vivo dosimetry is one of the quality assurance tools used in radiotherapy to monitor the dose delivered to the patient. Electronic portal imaging device (EPID) images for a set of solid water phantoms of varying thicknesses were acquired and the data fitted onto a quadratic equation, which relates the reduction in photon beam intensity to the attenuation coefficient and material thickness at a reference condition. The quadratic model is used to convert the measured grey scale value into water equivalent path length (EPL) at each pixel for any material imaged by the detector. For any other non-reference conditions, scatter, field size and MU variation effects on the image were corrected by relative measurements using an ionization chamber and an EPID. The 2D EPL is linked to the percentage exit dose table, for different thicknesses and field sizes, thereby converting the plane pixel values at each point into a 2D dose map. The off-axis ratio is corrected using envelope and boundary profiles generated from the treatment planning system (TPS). The method requires field size, monitor unit and source-to-surface distance (SSD) as clinical input parameters to predict the exit dose, which is then used to determine the entrance dose. The measured pixel dose maps were compared with calculated doses from TPS for both entrance and exit depth of phantom. The gamma index at 3% dose difference (DD) and 3 mm distance to agreement (DTA) resulted in an average of 97% passing for the square fields of 5, 10, 15 and 20 cm. The exit dose EPID dose distributions predicted by the algorithm were in better agreement with TPS-calculated doses than phantom entrance dose distributions.

Citing Articles

An Empirical Transmitted EPID Dosimetry Method using a Back-Projection Algorithm.

S M H, M H B, M M, S N, S B, H G J Biomed Phys Eng. 2019; 9(5):551-558.

PMID: 31750269 PMC: 6820021. DOI: 10.31661/jbpe.v0i0.1082.


EPID based in vivo dosimetry system: clinical experience and results.

Celi S, Costa E, Wessels C, Mazal A, Fourquet A, Francois P J Appl Clin Med Phys. 2016; 17(3):262-276.

PMID: 27167283 PMC: 5690938. DOI: 10.1120/jacmp.v17i3.6070.


A dual two dimensional electronic portal imaging device transit dosimetry model based on an empirical quadratic formalism.

Tan Y, Metwaly M, Glegg M, Baggarley S, Elliott A Br J Radiol. 2015; 88(1051):20140645.

PMID: 25969867 PMC: 4628518. DOI: 10.1259/bjr.20140645.


Evaluation of six TPS algorithms in computing entrance and exit doses.

Tan Y, Metwaly M, Glegg M, Baggarley S, Elliott A J Appl Clin Med Phys. 2014; 15(3):229–240.

PMID: 24892349 PMC: 5711058. DOI: 10.1120/jacmp.v15i3.4739.


An empirical calibration method for an a-Si portal imaging device: applications in pretreatment verification of IMRT.

Conte L, Mordacchini C, Pozzi L, Vite C Radiol Med. 2012; 117(6):1044-56.

PMID: 22430680 DOI: 10.1007/s11547-012-0808-y.