Performance Assessment Techniques for Doppler Radar Physiological Sensors

Overview
Date 2009 Dec 8
PMID 19963863
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

This paper presents a technique for assessing the performance of continuous wave Doppler radar systems for physiological sensing. The technique includes an artificial target for testing physiological sensing radar systems with motion analogous to human heart movement and software algorithms leveraging the capabilities of this target to simply test radar system performance. The mechanical target provides simple to complex patterns of motion that are stable and repeatable. Details of radar system performance can be assessed and the effects of configuration changes that might not appear with a human target can be observed when using this mechanical target.