» Articles » PMID: 19938046

Magnetic Resonance Imaging in Real Time: Advances Using Radial FLASH

Overview
Date 2009 Nov 26
PMID 19938046
Citations 42
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To develop technical advances for real-time magnetic resonance imaging (MRI) that allow for improved image quality and high frame rates.

Materials And Methods: The approach is based on a combination of fast low-angle shot (FLASH) MRI sequences with radial data sampling and view sharing of successive acquisitions. Gridding reconstructions provide images free from streaking or motion artifacts and with a flexible trade-off between spatial and temporal resolution. Immediate image reconstruction and online display is accomplished with the use of an unmodified 3 T MRI system. For receive coils with a large number of elements this process is supported by a user-selectable channel compression that is based on a principal component analysis and performed during initial preparation scans.

Results: In preliminary applications to healthy volunteers, real-time radial FLASH MRI visualized continuous movements of the temporomandibular joint during voluntary opening and closing of the mouth at high spatial resolution (0.75 mm in-plane) and monitored cardiac functions at high temporal resolution (20 images per second) during free breathing and without synchronization to the electrocardiogram.

Conclusion: Real-time radial FLASH MRI emerges as a simple and versatile tool for a large range of clinical applications.

Citing Articles

High-resolution volumetric dynamic magnetic resonance imaging of the wrist using an 8-channel flexible receive coil.

Zi R, Abbas B, Wang B, Walczyk J, Brown R, Petchprapa C Skeletal Radiol. 2024; .

PMID: 39560769 DOI: 10.1007/s00256-024-04829-7.


Measurement of Scapholunate Joint Space Width on Real-Time MRI-A Feasibility Study.

Ehmig J, Lehmann K, Engel G, Kuck F, Lotz J, Aeffner S Diagnostics (Basel). 2024; 14(11).

PMID: 38893703 PMC: 11172194. DOI: 10.3390/diagnostics14111177.


Ultra-rapid, Free-breathing, Real-time Cardiac Cine MRI Using GRASP Amplified with View Sharing and KWIC Filtering.

Fan L, Hong K, Allen B, Paul R, Carr J, Zhang S Radiol Cardiothorac Imaging. 2024; 6(1):e230107.

PMID: 38358330 PMC: 10912880. DOI: 10.1148/ryct.230107.


Non-contrast free-breathing 2D CINE compressed SENSE T1-TFE cardiovascular MRI at 3T in sedated young children for assessment of congenital heart disease.

Ristow I, Hancken-Pauschinger C, Zhang S, Stark M, Kaul M, Rickers C PLoS One. 2024; 19(2):e0297314.

PMID: 38330070 PMC: 10852305. DOI: 10.1371/journal.pone.0297314.


The future of CMR: All-in-one vs. real-time CMR (Part 2).

Contijoch F, Rasche V, Seiberlich N, Peters D J Cardiovasc Magn Reson. 2024; 26(1):100998.

PMID: 38237901 PMC: 11211235. DOI: 10.1016/j.jocmr.2024.100998.