Comparison of a New Transcutaneous Bilirubinometer (Bilimed) with Serum Bilirubin Measurements in Preterm and Full-term Infants

Overview
Journal BMC Pediatr
Publisher Biomed Central
Specialty Pediatrics
Date 2009 Nov 14
PMID 19909530
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The gold standard to assess hyperbilirubinemia in neonates remains the serum bilirubin measurement. Unfortunately, this is invasive, painful, and costly. Bilimed, a new transcutaneous bilirubinometer, suggests more accuracy compared to the existing non-invasive bilirubinometers because of its new technology. It furthermore takes into account different skin colours. No contact with the skin is needed during measurement, no additional material costs occur. Our aim was to assess the agreement between the Bilimed and serum bilirubin in preterm and term infants of different skin colours.

Methods: The transcutaneous bilirubin measurements were performed on the infant's sternum and serum bilirubin was determined simultaneously. The agreement between both methods was assessed by Pearson's correlation and by Bland-Altman analysis.

Results: A total of 117 measurement cycles were performed in 99 term infants (group1), further 47 measurements in 38 preterm infants born between 34 - 36 6/7 gestational weeks (group 2), and finally 21 measurements in 13 preterm infants born between 28 - 33 6/7 gestational weeks (group 3). The mean deviation and variability (+/- 2SD) of the transcutaneous from serum bilirubin were: -14 (+/- 144) micromol/l; -0.82 (+/- 8.4) mg/dl in group 1, +16 (+/- 91) micromol/l;+0.93(+/- 5.3) mg/dl in group 2 and -8 (+/- 76) micromol/l; -0.47 (+/- 4.4) mg/dl in group 3. These limits of agreement are too wide to be acceptable in a clinical setting. Moreover, there was to be a trend towards less good agreement with increasing bilirubin values.

Conclusion: Despite its new technology the Bilimed has no advantages, and more specifically no better agreement not only in term and near-term Caucasian infants, but also in non-Caucasian and more premature infants.

Citing Articles

Skin color influences transcutaneous bilirubin measurements: a systematic in vitro evaluation.

Dam-Vervloet A, Morsink C, Krommendijk M, Nijholt I, van Straaten H, Poot L Pediatr Res. 2024; .

PMID: 38368497 DOI: 10.1038/s41390-024-03081-y.


Validity of BiliDx as a point-of-care bilirubin measurement device to diagnose and monitor neonatal jaundice at Muhimbili National Hospital, an observational study.

Clemence P, Moshiro R, Manji K BMC Pediatr. 2024; 24(1):114.

PMID: 38350890 PMC: 10863160. DOI: 10.1186/s12887-024-04565-w.


Transcutaneous bilirubinometry versus total serum bilirubin measurement for newborns.

Okwundu C, Olowoyeye A, Uthman O, Smith J, Wiysonge C, Bhutani V Cochrane Database Syst Rev. 2023; 5:CD012660.

PMID: 37158489 PMC: 10167941. DOI: 10.1002/14651858.CD012660.pub2.


The Impact of Phototherapy on the Accuracy of Transcutaneous Bilirubin Measurements in Neonates: Optimal Measurement Site and Timing.

Ho S, Lin Y, Chen C Diagnostics (Basel). 2021; 11(9).

PMID: 34574069 PMC: 8466921. DOI: 10.3390/diagnostics11091729.


On the dysfunctional hemoglobins and cyanosis connection: practical implications for the clinical detection and differentiation of methemoglobinemia and sulfhemoglobinemia.

Askew S, Baranoski G Biomed Opt Express. 2018; 9(7):3284-3305.

PMID: 29984098 PMC: 6033548. DOI: 10.1364/BOE.9.003284.


References
1.
Slusher T, Angyo I, Bode-Thomas F, Akor F, Pam S, Adetunji A . Transcutaneous bilirubin measurements and serum total bilirubin levels in indigenous African infants. Pediatrics. 2004; 113(6):1636-41. DOI: 10.1542/peds.113.6.1636. View

2.
Bhutani V, Gourley G, Adler S, Kreamer B, Dalin C, Johnson L . Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia. Pediatrics. 2000; 106(2):E17. DOI: 10.1542/peds.106.2.e17. View

3.
Tan K, Mylvaganam A . Transcutaneous bilirubinometry in preterm very low birthweight infants. Acta Paediatr Scand. 1988; 77(6):796-801. DOI: 10.1111/j.1651-2227.1988.tb10758.x. View

4.
Brown L, Arnold L, Allison D, Jacobsen B, Klein M, Charsha D . Transcutaneous bilirubinometer: intermeter reliability. J Perinatol. 1990; 10(2):167-9. View

5.
Jeffrey Maisels M, Ostrea Jr E, Touch S, Clune S, Cepeda E, Kring E . Evaluation of a new transcutaneous bilirubinometer. Pediatrics. 2004; 113(6):1628-35. DOI: 10.1542/peds.113.6.1628. View