» Articles » PMID: 19898840

Treatment of Elastase-induced Intracranial Aneurysms in New Zealand White Rabbits by Use of a Novel Neurovascular Embolization Stent Device

Overview
Journal Neuroradiology
Specialties Neurology
Radiology
Date 2009 Nov 10
PMID 19898840
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: This study aims to test a novel balloon expandable stent covered with a polytetrafluoroethylene membrane (neurovascular embolization cover (NEC), NFocus Neuromedical, Palo Alto, California) regarding angiographic and histologic aneurysm occlusion. Radiopacity, stent placement, navigation, flexibility, and intimal proliferation were also evaluated.

Methods: Eight aneurysms were induced in New Zealand white rabbits. Digital subtraction angiography (DSA) was performed directly after stent placement and after 5 and 10 min. Four and 8 weeks after stent placement, an intra-arterial DSA control was performed. The animals were then sacrificed and the aneurysms histologically evaluated.

Results: The radiopaque markers were clearly visible. Although all the stents were easily navigated into the subclavian artery, the limited flexibility of the stent resulted in straightening of the vessel in four cases. As a result, exact stent placement was achieved and acutely confirmed in only two cases. However, at sacrifice, angiographic and histologic occlusion was noted at follow-up in five aneurysms.

Conclusion: In tortuous anatomy, the relative stiffness of the stent makes exact stent placement challenging. This may have been exacerbated by the movement of the vessels due to proximity to the heart in this model. Future studies should evaluate whether existing residual flow into an aneurysm lumen might lead to embolization without any additional treatment. Anticoagulation remains a very important part of aneurysm treatment with stents. The trend toward aneurysm occlusion by excluding it from the blood circulation seems a promising method in future endovascular therapy. The NEC device shows good potential.

Citing Articles

Imaging of intracranial aneurysms in animals: a systematic review of modalities.

Cayron A, Morel S, Allemann E, Bijlenga P, Kwak B Neurosurg Rev. 2023; 46(1):56.

PMID: 36786880 PMC: 9928939. DOI: 10.1007/s10143-023-01953-1.


[Flow diverter: a new therapy option for intracranial aneurysms].

Simgen A, Junk D, Reith W Radiologe. 2012; 52(12):1118-24.

PMID: 23178787 DOI: 10.1007/s00117-012-2373-5.


Flat-detector computed tomography in the assessment of intracranial stents: comparison with multi detector CT and conventional angiography in a new animal model.

Struffert T, Ott S, Adamek E, Schwarz M, Engelhorn T, Kloska S Eur Radiol. 2011; 21(8):1779-87.

PMID: 21365196 DOI: 10.1007/s00330-011-2093-7.

References
1.
Grunwald I, Romeike B, Roth C, Struffert T, Eymann R, Reith W . Anticoagulation regimes and their influence on the occlusion rate of aneurysms: an experimental study in rabbits. Neurosurgery. 2005; 57(5):1048-55. DOI: 10.1227/01.neu.0000180037.43693.fe. View

2.
Chiaradio J, Guzman L, Padilla L, Chiaradio M . Intravascular graft stent treatment of a ruptured fusiform dissecting aneurysm of the intracranial vertebral artery: technical case report. Neurosurgery. 2002; 50(1):213-6; discussion 216-7. DOI: 10.1097/00006123-200201000-00034. View

3.
Yoshioka T, Wright K, Wallace S, Lawrence Jr D, GIANTURCO C . Self-expanding endovascular graft: an experimental study in dogs. AJR Am J Roentgenol. 1988; 151(4):673-6. DOI: 10.2214/ajr.151.4.673. View

4.
Islak C, Kocer N, Albayram S, Kizilkilic O, Uzma O, Cokyuksel O . Bare stent-graft technique: a new method of endoluminal vascular reconstruction for the treatment of giant and fusiform aneurysms. AJNR Am J Neuroradiol. 2002; 23(9):1589-95. PMC: 7976808. View

5.