Computer-navigated joint arthroplasty surgery using optical tracking systems requires arrays fixated to bone via pins. Reports of fractures at pin sites have raised concern about safety. We reviewed the postoperative complications occurring in a single-surgeon series of 984 consecutive primary total knee arthroplasties. All pins were placed unicortically and connected by a dual pin array. Femoral pins were placed into the medial epicondyle, and tibial pins were placed in the shaft 10 cm inferior to the joint line. There were no fractures. Seventeen (1.7%) patients had minor pin-related complications. Twelve patients had a superficial infection around the tibial pin sites, which resolved with antibiotics. None of the infections required readmission or reoperation. We believe pin placement to be safe and effective with proper technique.

Citing Articles

Total versus robotic-assisted unicompartmental knee replacement (TRAKER) for medial compartment osteorthritis: a randomized controlled trial.

Clement N, Scott C, Macpherson G, Simpson P, Leitch G, Patton J Bone Jt Open. 2025; 6(2):164-177.

PMID: 39923804 PMC: 11809069. DOI: 10.1302/2633-1462.62.BJO-2024-0155.R1.


A comparison of pin site complications between large and small pin diameters in robotic-assisted total knee arthroplasty.

Desai S, Kunes J, Held M, Ren M, deMeireles A, Geller J J Exp Orthop. 2023; 10(1):22.

PMID: 36897468 PMC: 10006377. DOI: 10.1186/s40634-023-00584-1.


Is sequential bilateral robotic total knee arthroplasty a safe procedure? A matched comparative pilot study.

Batailler C, Anderson M, Flecher X, Ollivier M, Parratte S Arch Orthop Trauma Surg. 2022; 143(3):1599-1609.

PMID: 35536354 DOI: 10.1007/s00402-022-04455-9.


Do bicortical diaphyseal array pins create the risk of periprosthetic fracture in robotic-assisted knee arthroplasties?.

Yun A, Qutami M, Pasko K Arthroplasty. 2022; 3(1):25.

PMID: 35236500 PMC: 8796534. DOI: 10.1186/s42836-021-00082-8.


Robotic- and orthosensor-assisted versus manual (ROAM) total knee replacement: a study protocol for a randomised controlled trial.

Clement N, Bardgett M, Galloway S, Baron Y, Smith K, Weir D Trials. 2022; 23(1):70.

PMID: 35065657 PMC: 8783439. DOI: 10.1186/s13063-021-05936-9.