Adjustable Shunt Valve-induced Magnetic Resonance Imaging Artifact: a Comparative Study

Overview
Journal J Neurosurg
Specialty Neurosurgery
Date 2009 Oct 13
PMID 19817540
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Object: In this paper, the authors' goal was to compare the artifact induced by implanted (in vivo) adjustable shunt valves in spin echo, diffusion weighted (DW), and gradient echo MR imaging pulse sequences.

Methods: The MR images obtained in 8 patients with proGAV and 6 patients with Strata II adjustable shunt valves were assessed for artifact areas in different planes as well as the total volume for different pulse sequences.

Results: Artifacts induced by the Strata II valve were significantly larger than those induced by proGAV valve in spin echo MR imaging pulse sequence (29,761 vs 2450 mm(3) on T2-weighted fast spin echo, p = 0.003) and DW images (100,138 vs 38,955 mm(3), p = 0.025). Artifacts were more marked on DW MR images than on spin echo pulse sequence for both valve types.

Conclusions: Adjustable valve-induced artifacts can conceal brain pathology on MR images. This should influence the choice of valve implantation site and the type of valve used. The effect of artifacts on DW images should be highlighted pending the development of less MR imaging artifact-inducing adjustable shunt valves.

Citing Articles

Field strength difference in extent of artifacts induced by CERTAS Plus valves in patients with idiopathic normal pressure hydrocephalus.

Camerucci E, Elder B, Shu Y, Messina S, Gunter J, Graff-Radford J Neuroradiol J. 2023; 36(6):665-673.

PMID: 37118867 PMC: 10649542. DOI: 10.1177/19714009231173099.


Programmable Shunt Valves for Pediatric Hydrocephalus: 22-Year Experience from a Singapore Children's Hospital.

Tey M, Ng L, Low D, Seow W, Low S Brain Sci. 2021; 11(11).

PMID: 34827547 PMC: 8615584. DOI: 10.3390/brainsci11111548.


Valve exchange towards an adjustable differential pressure valve with gravitational unit, clinical outcome of a single-center study.

Alavi S, Schulz M, Schaumann A, Schwarz K, Thomale U Childs Nerv Syst. 2017; 33(5):759-765.

PMID: 28332153 DOI: 10.1007/s00381-017-3387-0.


The Contribution of Common Surgically Implanted Hardware to Functional MR Imaging Artifacts.

Desai A, Strother M, Faraco C, Morgan V, Ladner T, Dethrage L AJNR Am J Neuroradiol. 2015; 36(11):2068-73.

PMID: 26272973 PMC: 4644697. DOI: 10.3174/ajnr.A4419.


3T magnetic resonance imaging testing of externally programmable shunt valves.

Zabramski J, Preul M, Debbins J, McCusker D Surg Neurol Int. 2012; 3:81.

PMID: 22937481 PMC: 3424682. DOI: 10.4103/2152-7806.99171.