Arthrodesis of the Equine Proximal Interphalangeal Joint: a Biomechanical Comparison of Two Parallel Headless, Tapered, Variable-pitched, Titanium Compression Screws and Two Parallel 5.5 Mm Stainless-steel Cortical Screws

Overview
Journal Vet Surg
Date 2009 Sep 29
PMID 19781029
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare the biomechanical characteristics, failure mode, and effects of side (left or right limb) and end (forelimb or hindlimb) of different screws in 2-screw, parallel-screw proximal interphalangeal joint arthrodesis constructs in horses.

Study Design: In vitro experimental study.

Sample Population: Twenty limbs from 6 cadavers (4 complete limb sets, 2 partial sets-total of 4 forelimb and 6 hindlimb pairs).

Methods: Two parallel 5.5 mm cortical (AO) screws were inserted in lag fashion in 1 randomly allocated limb of a pair, and 2 parallel headless, tapered, variable-pitched, titanium compression screws (Acutrak-Plus) were inserted in the contralateral limb. Constructs were tested in 3-point bending in a dorsopalmar (plantar) direction using a materials-testing machine at a loading rate of 5.83 mm/s. Maximal bending moment at failure and composite stiffness were calculated from data generated on load-displacement curves. Data were analyzed using a Friedman 2-way analysis of variance and Wilcoxon's signed-rank tests.

Results: No significant difference was detected for bending moment or stiffness values in proximal interphalangeal joint arthrodesis constructs using 2 parallel Acutrak-Plus or AO screws for fixation. Mean stiffness values were significantly different between forelimb and hindlimb constructs.

Conclusions: Performance of 2 parallel Acutrak-Plus screws was biomechanically comparable with 2 parallel AO 5.5 mm cortical screws in in vitro pastern arthrodesis constructs.

Clinical Relevance: Acutrak-Plus screws may provide an alternative means of fixation for proximal interphalangeal joint arthrodesis.

Citing Articles

The Comparison of Latero-Medial versus Dorso-Palmar/Plantar Drilling for Cartilage Removal in the Proximal Interphalangeal Joint.

Spadari A, Forni G, Magno S, Tagliavia C, Canova M, Grandis A Animals (Basel). 2021; 11(6).

PMID: 34205601 PMC: 8235712. DOI: 10.3390/ani11061838.


Comparison of limited-contact dynamic compression plate and locking compression plate constructs for proximal interphalangeal joint arthrodesis in the horse.

Rocconi R, Carmalt J, Sampson S, Elder S, Gilbert E Can Vet J. 2015; 56(6):615-9.

PMID: 26028685 PMC: 4431161.