Background: Prior studies show that lactate is a useful prognostic marker in sepsis.
Objective: To study the feasibility and accuracy of a point-of-care (POC) analyzer capable of performing bedside serum lactate measurements; and to determine if other measurements (pH, base excess) are predictive of mortality.
Methods:
Design: prospective cohort study of adult (age 18 years or older) Emergency Department (ED) patients with suspected infection during the study period of May 2006 through March 2007.
Setting: A 55,000-annual-visit urban tertiary care ED.
Intervention: A point-of-care device (i-STAT, Abbott Point of Care Inc., Princeton, NJ) was deployed using a standardized training and quality assurance process. Using POC testing, we measured serum lactate, pH, and base excess, as well as concomitant lactate measurement via a central laboratory.
Statistics: Area under the curve (AUC) for receiver operator characteristic curve, Bland-Altman statistics along with a correlation coefficient, and relative risk with 95% confidence intervals reported.
Results: There were 699 patients enrolled, of whom 34 (4.9%) died. The AUCs for mortality prediction were: POC lactate 0.72, laboratory lactate 0.70, pH measurement 0.60, and base excess 0.60. Bland-Altman showed that POC lactate was, on average, 0.32 (95% confidence interval -0.35-0.98) lower than laboratory lactate, with agreement kappa = 0.97.
Conclusions: A point-of-care testing device provides a reliable and feasible way to measure serum lactate at the bedside. The pH and base excess were less helpful.
Lee S, Song J, Lee S, Kim S, Han K, Lee S J Clin Med. 2024; 13(18).
PMID: 39336876 PMC: 11431886. DOI: 10.3390/jcm13185389.
Gosselin M, Mabire C, Pasquier M, Carron P, Hugli O, Ageron F Intern Emerg Med. 2022; 17(6):1803-1812.
PMID: 35678940 PMC: 9178320. DOI: 10.1007/s11739-022-03005-w.
Wang T, Hsu Y Diagnostics (Basel). 2021; 11(8).
PMID: 34441408 PMC: 8394551. DOI: 10.3390/diagnostics11081474.
Charoentanyarak S, Sawunyavisuth B, Deepai S, Sawanyawisuth K J Prim Care Community Health. 2021; 12:21501327211000233.
PMID: 33733925 PMC: 7983462. DOI: 10.1177/21501327211000233.
Validation of a point-of-care capillary lactate measuring device (Lactate Pro 2).
Raa A, Sunde G, Bolann B, Kvale R, Bjerkvig C, Eliassen H Scand J Trauma Resusc Emerg Med. 2020; 28(1):83.
PMID: 32811544 PMC: 7437027. DOI: 10.1186/s13049-020-00776-z.