Rapid Blood Separation is Superior to Fluoride for Preventing in Vitro Reductions in Measured Blood Glucose Concentration

Overview
Journal J Clin Pathol
Specialty Pathology
Date 2009 Jul 30
PMID 19638548
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: To determine whether tubes containing sodium fluoride negatively bias blood glucose concentration by directly comparing glucose concentrations in paired blood samples collected in tubes containing lithium heparin (Li-Heparin) and tubes containing sodium fluoride/potassium oxalate (NaF-KOx).

Methods: Paired blood samples from a group of patients (n = 1040) were collected in tubes containing Li-Heparin and tubes containing NaF-KOx at the same time. All Li-Heparin samples were centrifuged soon after collection and were kept cool in transport along with NaF-KOx samples, which were centrifuged at the receiving location after an average transport time of 4 h, but immediately before analysis. Glucose concentrations in the paired samples were determined simultaneously by an automated oxidase method.

Results: The mean glucose concentrations for NaF-KOx samples and Li-Heparin samples were 5.7 mmol/l and 6.1 mmol/l, respectively, with a mean difference of 0.39 mmol/l.

Conclusion: Rapid separation of heparinised blood is superior to fluoride alone for abrogating glycolytic effects on blood glucose measurements in the clinical laboratory.

Citing Articles

Efficacy and Clinical Utility of Sodium Fluoride Tubes Versus Serum Tubes for Blood Glucose Estimation.

Daimane D, Basu S Cureus. 2024; 16(10):e72641.

PMID: 39610588 PMC: 11604175. DOI: 10.7759/cureus.72641.


Evaluation of point-of-care capillary and venous blood glucose concentrations in hospitalized neonatal foals.

Dias Moreira A, Weng H, Hostnik L, Beasley E, Peek S, Munsterman A J Vet Emerg Crit Care (San Antonio). 2024; 34(6):570-578.

PMID: 39558467 PMC: 11650952. DOI: 10.1111/vec.13429.


Optimizing d-mannose and glyceraldehyde concentrations as glucose preservatives without clinically affecting biochemical test results.

Wiriyaprasit R, Moonla K, Apiratmateekul N, Chittamma A, Kost G, Treebuphachatsakul W Pract Lab Med. 2024; 39:e00388.

PMID: 38623088 PMC: 11017339. DOI: 10.1016/j.plabm.2024.e00388.


Comparative and stability study of glucose concentrations measured in both sodium fluoride and serum separator tubes.

Dibbasey M, Umukoro S, Bojang A Pract Lab Med. 2024; 39:e00360.

PMID: 38313813 PMC: 10832486. DOI: 10.1016/j.plabm.2024.e00360.


Alternative pre-analytic sample handling techniques for glucose measurement in the absence of fluoride tubes in low resource settings.

Nakanga W, Balungi P, Niwaha A, Shields B, Hughes P, Andrews R PLoS One. 2022; 17(2):e0264432.

PMID: 35226681 PMC: 8884485. DOI: 10.1371/journal.pone.0264432.