» Articles » PMID: 1962346

Waveform Pattern Recognition--a New Semiquantitative Method for Analysis of Fetal Aortic and Umbilical Artery Blood Flow Velocity Recorded by Doppler Ultrasound

Overview
Specialty Radiology
Date 1991 Jan 1
PMID 1962346
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

A semiquantitative computerized waveform pattern recognition system for analysis of the fetal descending aortic and umbilical artery Doppler flow velocity waveforms is presented. Based on empirically and manually selected clinical recordings from both vessels, 11 computerized and normalized standard curves for the aorta (type curves A to K), and 10 curves for the umbilical artery (type curves a to j) were constructed. The best match between the normalized waveform and the standard curve was based on either the degree of absent diastolic flow or, in cases with positive diastolic flow, on the calculation of the least square sum of the difference. The pattern recognition was tested against conventional waveform indices and our older semiquantitative Blood Flow Class (BFC) system in 472 clinical consecutive Doppler recordings. A good correlation was found. This new relatively simple computer-based method for waveform analysis is now prospectively applied in clinical studies.

Citing Articles

Fetal Aortic and Umbilical Doppler Flow Velocity Waveforms in Pregnancy: The Concept of Aortoumbilical Column.

De Almeida A, Morais A, Miguel F, Gaio A, Guedes-Martins L Curr Cardiol Rev. 2024; .

PMID: 38441054 PMC: 11071678. DOI: 10.2174/011573403X255256230919061018.