Macular Thickness Measurements in Normal Eyes with Time-domain and Fourier-domain Optical Coherence Tomography

Overview
Journal Retina
Date 2009 Jul 9
PMID 19584656
Citations 26
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare macular thickness measurements using time-domain optical coherence tomography (OCT) and Fourier-domain OCT.

Methods: Thirty-two eyes from 32 normal patients underwent complete ophthalmic evaluation. Macular scanning using the StratusOCT and the RTVue-100 OCT were performed for a total of three times each on the same visit. The average retinal thicknesses of the nine macular sectors as defined by the Early Treatment Diabetic Retinopathy Study, along with the foveal center point and macular volume, were recorded. The SD, the coefficient of variation, and the intraclass correlation coefficient were calculated for each parameter studied. Comparisons were made between the two OCTs in terms of retinal thicknesses measurements, their reproducibility, and macular regional differences. Correlations between retinal thickness and demographic variables (age and gender) were also investigated. Due to known differences in segmentation algorithms of the two OCTs, software calipers were used to measure the distance from the internal limiting membrane to the photoreceptor inner segment-outer segment junction at the foveal center point on all RTVue scans to allow a more fair comparison.

Results: The RTVue yielded greater retinal thickness measurements in nearly all macular subfields compared with the StratusOCT. Even after accounting for differences in segmentation algorithms, significant disparities were still evident with the RTVue measurements less than those of the StratusOCT at the foveal center. On both machines, the macula was thinnest at the fovea and thickest within the 3 mm ring. There were some consistent regional variations in macular thickness evident on both OCTs. Compared with the StratusOCT, the RTVue generally had lower coefficients of variation and higher intraclass coefficients, suggesting better reproducibility. Age and gender also appeared to be important determinants in some macular thickness parameters.

Conclusion: Compared with StratusOCT, the RTVue fourier-domain OCT yields greater retinal thickness measurements with greater reproducibility, presumably due to different segmentation algorithms, increased sampling density, and greater resolution. However, regional differences across the macula can be consistently observed with both devices.

Citing Articles

Macular Thickness and Volume by Spectral-Domain Optical Coherence Tomography and their Related Factors in the Elderly Population.

Hashemi H, Heydarian S, Hashemi A, Khabazkhoob M J Curr Ophthalmol. 2024; 35(4):362-368.

PMID: 39281392 PMC: 11392310. DOI: 10.4103/joco.joco_153_23.


Factors influencing the reliability of measurements in eyes with full-thickness macular holes: are we measuring incorrectly?.

Moussa G, Jalil A, Lippera M, Alnafisee N, Ivanova T BMJ Open Ophthalmol. 2024; 9(1).

PMID: 39153755 PMC: 11331849. DOI: 10.1136/bmjophth-2023-001531.


Long-term Follow-up Results of Idiopathic Epiretinal Membrane Removal in the Eyes with Ectopic Inner Foveal Retinal Layers.

Tokuc E, Karabas L, Muftuoglu I, Seyyar S Middle East Afr J Ophthalmol. 2024; 30(2):72-79.

PMID: 39006936 PMC: 11238928. DOI: 10.4103/meajo.meajo_51_24.


Retinal and Choroidal Thickness in an Indigenous Population from Ghana: Comparison with Individuals with European or African Ancestry.

Zouache M, Faust C, Silvestri V, Akafo S, Lartey S, Mehta R Ophthalmol Sci. 2023; 4(2):100386.

PMID: 37868802 PMC: 10585639. DOI: 10.1016/j.xops.2023.100386.


Predictors of treatment outcomes following treat-and-extend regimen with aflibercept for branch retinal vein occlusion: post-hoc analysis of the PLATON trial.

Son W, Jeong W, Park J, Kim J, Ji Y, Sagong M Sci Rep. 2023; 13(1):11730.

PMID: 37474669 PMC: 10359445. DOI: 10.1038/s41598-023-38955-4.


References
1.
Wojtkowski M, Srinivasan V, Fujimoto J, Ko T, Schuman J, Kowalczyk A . Three-dimensional retinal imaging with high-speed ultrahigh-resolution optical coherence tomography. Ophthalmology. 2005; 112(10):1734-46. PMC: 1939719. DOI: 10.1016/j.ophtha.2005.05.023. View

2.
Chamberlain M, Guymer R, Dirani M, Hopper J, Baird P . Heritability of macular thickness determined by optical coherence tomography. Invest Ophthalmol Vis Sci. 2005; 47(1):336-40. DOI: 10.1167/iovs.05-0599. View

3.
Drexler W, Morgner U, Ghanta R, Kartner F, Schuman J, Fujimoto J . Ultrahigh-resolution ophthalmic optical coherence tomography. Nat Med. 2001; 7(4):502-7. PMC: 1950821. DOI: 10.1038/86589. View

4.
Pierre-Kahn V, Tadayoni R, Haouchine B, Massin P, Gaudric A . Comparison of optical coherence tomography models OCT1 and Stratus OCT for macular retinal thickness measurement. Br J Ophthalmol. 2005; 89(12):1581-5. PMC: 1772971. DOI: 10.1136/bjo.2005.069815. View

5.
Fujimoto J . Optical coherence tomography for ultrahigh resolution in vivo imaging. Nat Biotechnol. 2003; 21(11):1361-7. DOI: 10.1038/nbt892. View