» Articles » PMID: 19577211

Feasibility of Dual-source Cardiac CT Angiography with High-pitch Scan Protocols

Overview
Publisher Elsevier
Date 2009 Jul 7
PMID 19577211
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Cardiac CT angiography (CCTA) has become a frequently used diagnostic tool in clinical practice, but concern remains about the radiation exposure. Because of the second x-ray acquisition system, dual-source CT systems might allow for high-pitch CT data acquisition and thus for examination of the whole heart during a single heart beat, with the potential for radiation dose reduction.

Objective: We assessed the feasibility of a high-pitch scan mode with a dual-source CT system.

Methods: High-pitch modes were used in patients undergoing CCTA with a dual-source CT system. Diagnostic image quality for cardiac structures and coronary arteries was assessed. Radiation dose was estimated from the scanner-generated dose-length product (DLP).

Results: CCTA was performed in 14 patients during a single heart beat applying a pitch value of 3.4. Mean heart rate during examination was 56.4+/-8.1 beats/min. Diagnostic image quality for the assessment of larger cardiac structures was obtained in all patients, whereas diagnostic image quality could be achieved in 82% of all coronary segments. With a mean DLP of 145+/-47 mGy x cm, the resulting estimated radiation dose was 2.0+/-0.7 mSv.

Conclusions: This proof-of-concept study shows the ability of dual-source CT scanners to scan the whole heart during one single heart beat at low radiation dose.

Citing Articles

Optimizing coronary CT angiography quality with motion-compensated reconstruction for second-generation dual-layer spectral detector CT.

Liu S, Huang L, Gui S, Pang X, Liu J, Li X Eur Radiol. 2024; 35(1):381-392.

PMID: 38987398 DOI: 10.1007/s00330-024-10908-z.


Influence of heart rate and heart rate variability on the feasibility of ultra-fast, high-pitch coronary photon-counting computed tomography angiography.

Rotkopf L, Froelich M, Riffel P, Ziener C, Reid C, Schlemmer H Int J Cardiovasc Imaging. 2023; 39(5):1065-1073.

PMID: 36773035 PMC: 10160151. DOI: 10.1007/s10554-023-02808-y.


Contrast agent volume in coronary computer tomography angiography-where are the limits?.

Korosoglou G, Giusca S Quant Imaging Med Surg. 2021; 11(10):4511-4513.

PMID: 34604006 PMC: 8408785. DOI: 10.21037/qims-21-488.


Coronary Computer Tomography Angiography in 2021-Acquisition Protocols, Tips and Tricks and Heading beyond the Possible.

Giusca S, Schutz M, Kronbach F, Wolf D, Nunninger P, Korosoglou G Diagnostics (Basel). 2021; 11(6).

PMID: 34200866 PMC: 8230532. DOI: 10.3390/diagnostics11061072.


Exclusion of left atrial thrombus by dual-source cardiac computed tomography prior to catheter ablation for atrial fibrillation.

Kottmaier M, Jilek C, Berglar S, Reents T, Bourier F, Semmler V Clin Res Cardiol. 2018; 108(2):150-156.

PMID: 30051177 DOI: 10.1007/s00392-018-1333-0.