» Articles » PMID: 19571601

Comparison of Three Different Optical Coherence Tomography Models for Total Macular Thickness Measurements in Healthy Controls

Overview
Journal Ophthalmologica
Publisher Karger
Specialty Ophthalmology
Date 2009 Jul 3
PMID 19571601
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Background/aims: Conventional time domain optical coherence tomography (OCT) has become an important diagnostic tool to measure retinal thickness in clinical routine. Recently, different models of high-speed, high-resolution frequency domain OCTs have been introduced by various manufacturers. The purpose of this study was to compare 3 commercially available OCT models for retinal thickness measurements in healthy controls.

Methods: OCT scans were performed on 28 healthy eyes with the RTVue-100 FD-OCT (Optovue Inc., USA), the Cirrus HD-OCT (Carl Zeiss Meditec Inc., USA) and the Stratus OCT 3000 (Carl Zeiss Meditec Inc.). Retinal thickness values were calculated and compared between OCT models.

Results: Differences in mean retinal thickness measurements between OCT models were statistically significant. Mean retinal thickness measurement with Cirrus OCT, RTVue OCT and with Stratus OCT was 300, 265 and 257 microm, respectively.

Conclusion: Measurements with different OCT models lead to significantly different retinal thickness values.

Citing Articles

Association of retinal neurodegeneration with the progression of cognitive decline in Parkinson's disease.

Murueta-Goyena A, Romero-Bascones D, Teijeira-Portas S, Urcola J, Ruiz-Martinez J, Del Pino R NPJ Parkinsons Dis. 2024; 10(1):26.

PMID: 38263165 PMC: 10805713. DOI: 10.1038/s41531-024-00637-x.


Repeatability and Agreement of Macular Thickness Measurements Obtained with Two Different Scan Modes of the Optovue RTVue Optical Coherence Tomography Device.

Kola M, Onal M, Turk A, Erdol H Turk J Ophthalmol. 2019; 49(2):78-83.

PMID: 31055892 PMC: 6517858. DOI: 10.4274/tjo.galenos.2018.88972.


Assessment of Macular Thickness in Healthy Eyes Using Cirrus HD-OCT: A Cross-Sectional Study.

Sabouri M, Kazemnezhad E, Hafezi V Med Hypothesis Discov Innov Ophthalmol. 2017; 5(3):104-111.

PMID: 28293656 PMC: 5347212.


Macular thickness and macular volume measurements using spectral domain optical coherence tomography in normal Nepalese eyes.

Pokharel A, Shankar Shrestha G, Shrestha J Clin Ophthalmol. 2016; 10:511-9.

PMID: 27041990 PMC: 4811214. DOI: 10.2147/OPTH.S95956.


Evaluation of macular thickness change after inferior oblique muscle recession surgery.

Turan-Vural E, Unlu C, Erdogan G, Aykut A, Bayramlar H, Atmaca F Indian J Ophthalmol. 2014; 62(6):715-8.

PMID: 25005201 PMC: 4131326. DOI: 10.4103/0301-4738.136230.