Field Evaluation of Diagnostic Accuracy of an Oral Fluid Rapid Test for HIV, Tested at Point-of-service Sites in Rural Zimbabwe

Overview
Date 2009 Jun 18
PMID 19530953
Citations 27
Authors
Affiliations
Soon will be listed here.
Abstract

The objective of this study was to validate the use of OraQuick ADVANCE Rapid HIV-1/2 Antibody test (OraSure Technologies Inc., Bethlehem, PA) on oral fluid for a population-based HIV prevalence survey of rural youth in southeast Zimbabwe. The evaluation was conducted in patients presenting for voluntary counseling and testing at rural clinics. Each participant provided an oral fluid sample tested using OraQuick ADVANCE. In addition, dried blood specimens were collected and tested blind at the National Microbiology Reference Laboratory in Harare using two enzyme-linked immunosorbent assays (ELISA; Vironostika, Biomérieux BV, Boxtel, The Netherlands and Ani Labsystems, Ltd., Vantaa, Finland) with confirmatory Western blot (MP Diagnostics [formerly Genelabs Diagnostics], Medical Technology Promedt Consulting GMBH, St. Ingbert, Germany) for samples with discrepant results. Diagnostic accuracy of the oral fluid assay was determined against the ELISA/Western blot algorithm as gold standard. Five hundred and ninety-one participants took part in the study between February and July 2006. Sensitivity of the test on oral fluid was 100% (95% confidence interval [CI]: 97.9-100), and specificity was 100% (95% CI: 99.1-100). HIV prevalence based on the reference standard was 29.8% (95% CI: 26.1-33.5). This is one of the first validations of this rapid assay on oral fluid conducted in a general population to be reported in Africa. While there are some limitations with the assay (e.g., unlikely to detect those in early stages of HIV infection or with reduced viral load; altered accuracy in pregnancy) these limitations also apply to other rapid assays. The results showed the assay to be 100% accurate in determining HIV status, performed well in field settings, and can be considered suitable for use in epidemiologic surveys aiming to estimate HIV prevalence in general populations.

Citing Articles

A handheld HIV detection platform using paper-based sample preparation and real-time isothermal amplification.

Adedokun G, Sidhu G, Alipanah M, Wang G, Fan Z Microsyst Nanoeng. 2024; 10(1):181.

PMID: 39613739 PMC: 11607462. DOI: 10.1038/s41378-024-00822-1.


Preferences for oral-fluid-based or blood-based HIV self-testing and provider-delivered testing: an observational study among different populations in Zimbabwe.

Mavhu W, Makamba M, Hatzold K, Maringwa G, Takaruza A, Mutseta M BMC Infect Dis. 2023; 22(Suppl 1):973.

PMID: 37848810 PMC: 10583299. DOI: 10.1186/s12879-023-08624-y.


Validation study of an automated chemiluminiscence assay to detect HIV antibodies in oral fluid specimens.

Gonzalez V, Lopez N, Grifols J, Egea L, Rivaya B, Wang J Eur J Clin Microbiol Infect Dis. 2022; 41(6):907-911.

PMID: 35524829 PMC: 9078087. DOI: 10.1007/s10096-022-04450-3.


SARS Antibody Testing in Children: Development of Oral Fluid Assays for IgG Measurements.

Hoschler K, Ijaz S, Andrews N, Ho S, Dicks S, Jegatheesan K Microbiol Spectr. 2022; 10(1):e0078621.

PMID: 34985331 PMC: 8729769. DOI: 10.1128/spectrum.00786-21.


Shifting Coronavirus Disease 2019 Testing Policy and Research to Include the Full Translation Pipeline.

Catania J, Martin J, Dolcini M, Orellana E, Henne J Open Forum Infect Dis. 2021; 8(2):ofaa649.

PMID: 33575425 PMC: 7863862. DOI: 10.1093/ofid/ofaa649.


References
1.
Akanmu A, Akinsete I, Adesemoye E, Okany C . Evaluation of saliva-based diagnostic test kit for routine detection of antibodies to HIV. Afr J Med Med Sci. 2003; 30(4):305-8. View

2.
OConnell R, Merritt T, Malia J, VanCott T, Dolan M, Zahwa H . Performance of the OraQuick rapid antibody test for diagnosis of human immunodeficiency virus type 1 infection in patients with various levels of exposure to highly active antiretroviral therapy. J Clin Microbiol. 2003; 41(5):2153-5. PMC: 154669. DOI: 10.1128/JCM.41.5.2153-2155.2003. View

3.
Holm-Hansen C, Nyombi B, Nyindo M . Saliva-based HIV testing among secondary school students in Tanzania using the OraQuick rapid HIV1/2 antibody assay. Ann N Y Acad Sci. 2007; 1098:461-6. DOI: 10.1196/annals.1384.036. View

4.
Wesolowski L, MacKellar D, Facente S, Dowling T, Ethridge S, Zhu J . Post-marketing surveillance of OraQuick whole blood and oral fluid rapid HIV testing. AIDS. 2006; 20(12):1661-6. DOI: 10.1097/01.aids.0000238413.13442.ed. View

5.
Debattista J, Bryson G, Roudenko N, Dwyer J, Kelly M, Hogan P . Pilot of non-invasive (oral fluid) testing for HIV within a clinical setting. Sex Health. 2007; 4(2):105-9. DOI: 10.1071/sh07014. View