» Articles » PMID: 19488827

Development and Evaluation of a New Fully Automatic Motion Detection and Correction Technique in Cardiac SPECT Imaging

Overview
Journal J Nucl Cardiol
Date 2009 Jun 3
PMID 19488827
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Background: In cardiac SPECT perfusion imaging, motion correction of the data is critical to the minimization of motion introduced artifacts in the reconstructed images. Software-based (data-driven) motion correction techniques are the most convenient and economical approaches to fulfill this purpose. However, the accuracy is significantly affected by how the data complexities, such as activity overlap, non-uniform tissue attenuation, and noise are handled.

Methods: We developed STASYS, a new, fully automatic technique, for motion detection and correction in cardiac SPECT. We evaluated the performance of STASYS by comparing its effectiveness of motion correcting patient studies with the current industry standard software (Cedars-Sinai MoCo) through blind readings by two readers independently.

Results: For 204 patient studies from multiple clinical sites, the first reader identified (1) 69 studies with medium to large axial motion, of which STASYS perfectly or significantly corrected 86.9% and MoCo 72.5%; and (2) 20 studies with medium to large lateral motion, of which STASYS perfectly or significantly corrected 80.0% and MoCo 60.0%. The second reader identified (1) 84 studies with medium to large axial motion, of which STASYS perfectly or significantly corrected 82.2% and MoCo 76.2%; and (2) 34 studies with medium to large lateral motion, of which STASYS perfectly or significantly corrected 58.9% and MoCo 50.0%.

Conclusions: We developed a fully automatic software-based motion correction technique, STASYS, for cardiac SPECT. Clinical studies showed that STASYS was effective and corrected a larger percent of cardiac SPECT studies than the current industrial standard software.

Citing Articles

Evaluation of Rigid-Body Motion Compensation in Cardiac Perfusion SPECT Employing Polar-Map Quantification.

Pretorius P, Johnson K, King M IEEE Trans Nucl Sci. 2017; 63(3):1419-1425.

PMID: 28042170 PMC: 5193381. DOI: 10.1109/TNS.2016.2545407.


Gated SPECT myocardial perfusion imaging quality assurance in current and future practice.

Nichols K, Van Tosh A J Nucl Cardiol. 2016; 24(2):543-545.

PMID: 28004313 DOI: 10.1007/s12350-016-0752-4.


The effect of patient anxiety and depression on motion during myocardial perfusion SPECT imaging.

Lyra V, Kallergi M, Rizos E, Lamprakopoulos G, Chatziioannou S BMC Med Imaging. 2016; 16(1):49.

PMID: 27550022 PMC: 4994279. DOI: 10.1186/s12880-016-0153-9.


Motion detection and amelioration in a dedicated cardiac solid-state CZT SPECT device.

Kennedy J, Strauss H Med Biol Eng Comput. 2016; 55(4):663-671.

PMID: 27417733 DOI: 10.1007/s11517-016-1548-z.


Value of automatic patient motion detection and correction in myocardial perfusion imaging using a CZT-based SPECT camera.

van Dijk J, van Dalen J, Mouden M, Ottervanger J, Knollema S, Slump C J Nucl Cardiol. 2016; 25(2):419-428.

PMID: 27406376 PMC: 5869883. DOI: 10.1007/s12350-016-0571-7.


References
1.
Eisner R, Noever T, Nowak D, CARLSON W, Dunn D, Oates J . Use of cross-correlation function to detect patient motion during SPECT imaging. J Nucl Med. 1987; 28(1):97-101. View

2.
Friedman J, Van Train K, Maddahi J, Rozanski A, Prigent F, Bietendorf J . "Upward creep" of the heart: a frequent source of false-positive reversible defects during thallium-201 stress-redistribution SPECT. J Nucl Med. 1989; 30(10):1718-22. View

3.
Cooper J, NEUMANN P, McCandless B . Effect of patient motion on tomographic myocardial perfusion imaging. J Nucl Med. 1992; 33(8):1566-71. View

4.
Matsumoto N, Berman D, Kavanagh P, Gerlach J, Hayes S, Lewin H . Quantitative assessment of motion artifacts and validation of a new motion-correction program for myocardial perfusion SPECT. J Nucl Med. 2001; 42(5):687-94. View

5.
Geckle W, Frank T, Links J, Becker L . Correction for patient and organ movement in SPECT: application to exercise thallium-201 cardiac imaging. J Nucl Med. 1988; 29(4):441-50. View