AcrySof Phakic Angle-supported Intraocular Lens for the Correction of Moderate-to-high Myopia: One-year Results of a Multicenter European Study
Overview
Authors
Affiliations
Purpose: To investigate the safety and effectiveness of the AcrySof phakic angle-supported intraocular lens (IOL) (Alcon Laboratories, Inc., Fort Worth, TX) for correction of moderate-to-high myopia in adults.
Design: One-year interim analysis of a phase 3, nonrandomized, open-label, prospective, multicenter European clinical study.
Participants: A total of 190 subjects (190 eyes) with moderate-to-high myopia. The preoperative mean manifest refraction spherical equivalent (MRSE) was -10.38 diopters (D) +/-2.43 standard deviation (SD).
Methods: Unilateral implantation of the AcrySof phakic angle-supported IOL.
Main Outcome Measures: Best spectacle-corrected visual acuity (BSCVA), uncorrected distance visual acuity (UCVA), predictability and stability of MRSE, adverse events, and endothelial cell density.
Results: Of 190 subjects enrolled, 161 completed the 1-year postoperative visit. No subjects lost > or =2 lines BSCVA. A UCVA of 20/20 or better was achieved by 57.8%; 99.4% had 20/40 or better. A BSCVA of 20/32 or better was achieved by 100% of subjects; 85.7% had 20/20 or better. The mean MRSE was -0.23 D (+/-0.50 D: -2.50 to 0.75 D). Residual refractive error was within +/-1.0 D from the target for 95.7% of subjects and within +/-0.5 D for 72.7% of subjects. The overall mean percentage change in central endothelial cell density 1 year after surgery was -4.77+/-8.04% (n = 139). No pupil ovalization, pupillary block, or retinal detachment events were observed.
Conclusions: The AcrySof phakic angle-supported IOL yielded excellent refractive correction and predictability with acceptable safety in subjects with moderate-to-high myopia. These 1-year interim analysis findings demonstrate preliminary support for the safety and efficacy of this IOL.
Musayeva A, Riedl J, Gericke A, Vossmerbaeumer U J Ophthalmol. 2022; 2022:5362020.
PMID: 35378887 PMC: 8976641. DOI: 10.1155/2022/5362020.
Phakic intraocular lenses: An overview.
Jonker S, Berendschot T, Saelens I, Bauer N, Nuijts R Indian J Ophthalmol. 2020; 68(12):2779-2796.
PMID: 33229653 PMC: 7856940. DOI: 10.4103/ijo.IJO_2995_20.
Al-Qahtani S, Al-Afraj K, Al-Jindan M, Al-Beshri A, Khandekar R Int J Ophthalmol. 2018; 11(5):888-890.
PMID: 29862192 PMC: 5957045. DOI: 10.18240/ijo.2018.05.27.
Kohnen T, Shajari M Ophthalmologe. 2016; 113(6):529-38.
PMID: 27277751 DOI: 10.1007/s00347-016-0283-z.
[Scheimpflug photography for the examination of phakic intraocular lenses].
Baumeister M Ophthalmologe. 2014; 111(10):935-41.
PMID: 25332044 DOI: 10.1007/s00347-013-2963-2.