» Articles » PMID: 19324853

Test-retest Variability of Microperimetry Using the Nidek MP1 in Patients with Macular Disease

Overview
Specialty Ophthalmology
Date 2009 Mar 28
PMID 19324853
Citations 59
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To determine the test-retest variability of the retinal sensitivity of the Nidek MP1 microperimeter in patients with macular disease.

Methods: In this prospective study, 50 patients were enrolled with a range of macular diseases. One examiner performed two consecutive microperimetry tests for all patients using the same test strategy. Test-retest variability for mean sensitivity (MS), mean deviation (MD), point-wise sensitivity (PWS), local defect classification (LDC), average sensitivity for the central macula (CMS, 16 loci inside 10 degrees ), paracentral macular sensitivity (PMS, 52 loci in the 10 to 20 degrees ring), and dense scotoma size (DSS) were analyzed by calculating the 95% coefficients of repeatability or percentage agreement.

Results: Mean (SD) age and visual acuity were 61 (15) years and 0.34 (0.32) logMAR, respectively. The mean difference in MS between tests 1 and 2 was +0.2 dB (SD, 0.9 dB; P = 0.127). The coefficients of repeatability for MS, MD, CMS, and PMS were 1.81, 2.56, 2.13, and 1.93 dB, respectively. The mean (SD) of coefficients of repeatability for PWS across all 68 loci was 5.56 (0.86) dB. Of all test loci in all patients 76% had perfect agreement in LDC, and 94% of patients had a change in DSS of four or fewer test loci.

Conclusions: Test-retest variability was lowest for MS and highest for PWS. However, MS does not provide spatial information. The authors recommend the use of CMS and PMS for monitoring macular function and consider a change of greater than 2.56 and 2.31 dB (the upper limit of the 95% confidence interval of their coefficients of repeatability), respectively, to exceed test-retest variability.

Citing Articles

Photoreceptor Function and Structure in Retinal Areas With Intraretinal Hyperreflective Foci in Age-Related Macular Degeneration.

Wang X, Hoshi S, Liu R, Corradetti G, Ip M, Sarraf D Invest Ophthalmol Vis Sci. 2025; 66(2):27.

PMID: 39928312 PMC: 11812613. DOI: 10.1167/iovs.66.2.27.


Longitudinal assessment of female carriers of choroideremia using multimodal retinal imaging.

Gocuk S, Ayton L, Edwards T, McGuinness M, MacLaren R, Taylor L Br J Ophthalmol. 2024; 109(2):293-299.

PMID: 39122355 PMC: 11866302. DOI: 10.1136/bjo-2024-325578.


Visual Field Tests: A Narrative Review of Different Perimetric Methods.

Rai B, Sabeti F, Carle C, Maddess T J Clin Med. 2024; 13(9).

PMID: 38730989 PMC: 11084906. DOI: 10.3390/jcm13092458.


Microperimetric evaluation and predictive factors of visual recovery after successful inverted internal limiting membrane-flap technique for macular hole in high myopic eyes.

Sborgia A, Boscia G, Niro A, Landini L, Pastore V, Albano V Front Med (Lausanne). 2023; 10:1276502.

PMID: 38076261 PMC: 10702352. DOI: 10.3389/fmed.2023.1276502.


Investigating the Spatiotemporal Summation of Perimetric Stimuli in Dry Age-Related Macular Degeneration.

Hunter A, Anderson R, Redmond T, Garway-Heath D, Mulholland P Transl Vis Sci Technol. 2023; 12(11):37.

PMID: 38019498 PMC: 10691387. DOI: 10.1167/tvst.12.11.37.