Pressure-frequency Sensing Subxiphoid Access System for Use in Percutaneous Cardiac Electrophysiology: Prototype Design and Pilot Study Results
Overview
Biophysics
Affiliations
We have designed, built, and tested an early prototype of a novel subxiphoid access system intended to facilitate epicardial electrophysiology, but with possible applications elsewhere in the body. The present version of the system consists of a commercially available insertion needle, a miniature pressure sensor and interconnect tubing, read-out electronics to monitor the pressures measured during the access procedure, and a host computer with user-interface software. The nominal resolution of the system is < 0.1 mmHg, and it has deviations from linearity of < 1%. During a pilot series of human clinical studies with this system, as well as in an auxiliary study done with an independent method, we observed that the pericardial space contained pressure-frequency components related to both the heart rate and respiratory rate, while the thorax contained components related only to the respiratory rate, a previously unobserved finding that could facilitate access to the pericardial space. We present and discuss the design principles, details of construction, and performance characteristics of this system.
Therapeutic payload delivery to the myocardium: Evolving strategies and obstacles.
Shazly T, Smith A, Uline M, Spinale F JTCVS Open. 2022; 10:185-194.
PMID: 36004211 PMC: 9390211. DOI: 10.1016/j.xjon.2022.04.043.
Mahapatra S, LaPar D, Bhamidipati C, McDaniel G, Kamath S, Bunch T Europace. 2011; 13(4):548-54.
PMID: 21296778 PMC: 3065918. DOI: 10.1093/europace/eur017.
Mahapatra S, Tucker-Schwartz J, Wiggins D, Gillies G, Mason P, McDaniel G Heart Rhythm. 2010; 7(5):604-9.
PMID: 20156612 PMC: 3047445. DOI: 10.1016/j.hrthm.2010.01.011.