Evaluation of a Computerized System for Mechanical Ventilation of Infants

Overview
Publisher Springer
Date 2009 Mar 6
PMID 19263230
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To evaluate a computerized system for mechanical ventilation of infants.

Methods: FLEX is a computerized system that includes the features of a patented mode known as adaptive-support ventilation (ASV). In addition, it has many other features including adjustment of positive end-expiratory pressure (PEEP), fraction of inspired oxygen (F(IO2)), minute ventilation, and control of weaning. It is used as an open-loop decision support system or as a closed-loop technique. Blood gas and ventilation data were collected from 12 infants in the neonatal intensive care at baseline and at the next round of evaluation. This data were input to open-loop version of FLEX. The system recommendations were compared to clinical determinations.

Results: FLEX recommended values for ventilation were on the average within 25% and 16.5% of the measured values at baseline and at the next round of evaluation, respectively. For F(IO2) and PEEP, FLEX recommended values were in general agreement with the clinical settings. FLEX recommendations for weaning were the same as the clinical determinations 50% of the time at baseline and 55% of the time at the next round of evaluation. FLEX did not recommend weaning for infants with weak spontaneous breathing effort or those who showed signs of dyspnea.

Conclusions: A computerized system for mechanical ventilation is evaluated for treatment of infants. The results of the study show that the system has good potential for use in neonatal ventilatory care. Further refinements can be made in the system for very low-birth-weight infants.

Citing Articles

Automated oxygen delivery for preterm infants with respiratory dysfunction.

Stafford I, Lai N, Tan K Cochrane Database Syst Rev. 2023; 11:CD013294.

PMID: 38032241 PMC: 10688253. DOI: 10.1002/14651858.CD013294.pub2.


A Pediatric Intensive Care Unit Bedside Computer Clinical Decision Support Protocol for Hyperglycemia Is Feasible, Safe and Offers Advantages.

Hirshberg E, Lanspa M, Wilson E, Sward K, Jephson A, Larsen G Diabetes Technol Ther. 2017; 19(3):188-193.

PMID: 28248127 PMC: 5359657. DOI: 10.1089/dia.2016.0423.


Computerized decision support in adult and pediatric critical care.

Williams C, Bratton S, Hirshberg E World J Crit Care Med. 2014; 2(4):21-8.

PMID: 24701413 PMC: 3953873. DOI: 10.5492/wjccm.v2.i4.21.


Feasibility and reliability of an automated controller of inspired oxygen concentration during mechanical ventilation.

Saihi K, Richard J, Gonin X, Kruger T, Dojat M, Brochard L Crit Care. 2014; 18(2):R35.

PMID: 24552490 PMC: 4031979. DOI: 10.1186/cc13734.


Dead space reduction by Kolobow's endotracheal tube does not justify the waiving of volume monitoring in small, ventilated lungs.

Proquitte H, Wendel R, Roehr C, Wauer R, Schmalisch G J Clin Monit Comput. 2014; 28(6):605-11.

PMID: 24469384 DOI: 10.1007/s10877-014-9559-5.


References
1.
Saura P, Blanch L, Mestre J, Valles J, Artigas A, Fernandez R . Clinical consequences of the implementation of a weaning protocol. Intensive Care Med. 1996; 22(10):1052-6. DOI: 10.1007/BF01699227. View

2.
Epstein S, Ciubotaru R . Independent effects of etiology of failure and time to reintubation on outcome for patients failing extubation. Am J Respir Crit Care Med. 1998; 158(2):489-93. DOI: 10.1164/ajrccm.158.2.9711045. View

3.
Moya F, Gadzinowski J, Bancalari E, Salinas V, Kopelman B, Bancalari A . A multicenter, randomized, masked, comparison trial of lucinactant, colfosceril palmitate, and beractant for the prevention of respiratory distress syndrome among very preterm infants. Pediatrics. 2005; 115(4):1018-29. DOI: 10.1542/peds.2004-2183. View

4.
Davis J, Notter R, Bhutani V, Kendig J, Shapiro D . Changes in pulmonary mechanics after the administration of surfactant to infants with respiratory distress syndrome. N Engl J Med. 1988; 319(8):476-9. DOI: 10.1056/NEJM198808253190804. View

5.
Esteban A, Frutos F, Tobin M, Alia I, Solsona J, Valverdu I . A comparison of four methods of weaning patients from mechanical ventilation. Spanish Lung Failure Collaborative Group. N Engl J Med. 1995; 332(6):345-50. DOI: 10.1056/NEJM199502093320601. View