Precision of Cortical Bone Reconstruction Based on 3D CT Scans

Overview
Specialty Radiology
Date 2009 Feb 17
PMID 19217257
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

The precision and accuracy of human cortical bone reconstruction using 3D CT scans was evaluated using machined bone segments. Both linear and angular errors were measured. Cadaver adult femoral and tibial cortical bone segments were obtained and machined in six orthogonal planes with a precision milling machine. CT scans were then obtained and the bone segments were reconstructed as digital replicas. Dimensional and angular measurements errors were evaluated for the machined bone segments and the results were compared with known dimensions based on milling machine settings to calculate errors due to scanning and model reconstruction. The model dimensional error in the coronal, sagittal and axial directions had a mean of 0.21 mm, with standard a deviation of 0.12 mm and a maximum error of 0.47 mm. The mean percent error was 0.74% and the maximum percent error was 1.9%. The angular error of models in the coronal, sagittal and axial directions was calculated, yielding a mean of 0.47 degrees with a standard deviation of 0.37 degrees and a maximum of 1.33 degrees. The error in the cross-sectional axial direction had a mean of 0.54 mm with a maximum error of 0.83 mm, depending on the slice interval. The main error source was of the image processing, which was about 70% of the total error. We found that machining cortical bone segments prior to CT scanning is an effective method for accuracy evaluation of CT-based bone reconstruction. This method can provide a reference for assessing the sensitivity, reliability and accuracy of CT-based applications in the study of movement, finite element modeling, and prosthesis construction.

Citing Articles

Computed tomography and structured light imaging guided orthopedic navigation puncture system: effective reduction of intraoperative image drift and mismatch.

He Z, Xu G, Zhang G, Wang Z, Sun J, Li W Front Surg. 2024; 11:1476245.

PMID: 39450295 PMC: 11499228. DOI: 10.3389/fsurg.2024.1476245.


A Practical Method for Slice Spacing Measurement Using the American Association of Physicists in Medicine Computed Tomography Performance Phantom.

Anam C, Naufal A, Dwihapsari Y, Fujibuchi T, Dougherty G J Med Phys. 2024; 49(1):103-109.

PMID: 38828077 PMC: 11141755. DOI: 10.4103/jmp.jmp_155_23.


A Comprehensive Review of Vision-Based 3D Reconstruction Methods.

Zhou L, Wu G, Zuo Y, Chen X, Hu H Sensors (Basel). 2024; 24(7).

PMID: 38610525 PMC: 11014007. DOI: 10.3390/s24072314.


Database of segmentations and surface models of bones of the entire lower body created from cadaver CT scans.

Fischer M Sci Data. 2023; 10(1):763.

PMID: 37923754 PMC: 10624917. DOI: 10.1038/s41597-023-02669-z.


Visualization and quantification of the degenerative pattern of the distal tibia and fibula in unilateral varus ankle osteoarthritis.

Seki H, Ogihara N, Kokubo T, Nagura T Sci Rep. 2021; 11(1):21628.

PMID: 34732780 PMC: 8566530. DOI: 10.1038/s41598-021-00874-7.